Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

Measurement area: Ø 10 mm
Acceptance angle: ± 5°
Accuracy guaranteed measurement distance: 30 ± 5 mm
Luminance:
Accuracy guaranteed range: 0.002 to 10,000 cd/m2
Accuracy (for white):
> 0.0001 cd/m2: -
> 0.0003 cd/m2: -
> 0.0005 cd/m2: -
> 0.001 cd/m2: ± 9 % (0.002 to cd/m2)
> 0.01 cd/m2: ± 2.5 %
> 0.1 cd/m2: ± 2 %
> 1 cd/m2: ± 2 %
> 10 cd/m2: ± 1.5 %
> 100 cd/m2: ± 1.5 %
Chromaticity:
Accuracy guaranteed luminance range: 0.01 to 10,000 cd/m2
Accuracy (for white):
> 0.001 cd/m2: -
> 0.003 cd/m2: -
> 0.01 cd/m2: ± 0.006
> 0.1 cd/m2: ± 0.002
> 1 cd/m2: ± 0.002
> 10 cd/m2: ± 0.002
> 100 cd/m2: ± 0.002
Parameters:
Lvxy (∆Lv ∆x ∆y)
Lvu’v’(∆Lv ∆u’ ∆v’)
Lv TCP duv (∆Lv ∆TCP duv)
X Y Z (∆X ∆Y ∆Z)
Lv λd Pe (∆Lv ∆λd ∆Pe)
Flicker (Contrast)
Flicker (JEITA)
Measurement synchronization mode:
Measurement speed mode: AUTO, LTD. AUTO, SLOW, FAST
Measurement target (Vertical synchronization frequency): 0.5 to 240 Hz (luminance and chromaticity), 0.5 to 130 Hz (flicker)
User calibration memory channel: 99 channels
Interface:
Communication: USB2.0, RS-232C
Trigger: IN: 1.8 V /3.3 to 5 V switching Out: 5 V
Power supply: DC 5 V (input from USB bus power line or RS communication connector)
Operation temperature/humidity range: 10 to 35°C, relative humidity 85% or less with no condensation
Storage temperature/humidity range: 0 to 45°C, relative humidity 85% or less (at 35°C) with no condensation