Measurement: X-ray fluorescence analysis
Determination method: Energy dispersion type
Determination of the object: Solid, liquid, powder
Range of measurement: 11Na-92U
Equipment size: W570mm x D402mm x H400mm
Equipment weight: 47kg
Sample room size: W368mm x D304mm x H78mm
Openway: The automatic control
X - ray generating department
X-ray tube: High power side window X-ray tube
Target materials: Target W (standard), Au, Ag, Rh (optional)
Voltage: 5-50KV
Current: 0-1000μA
Cooling way: Air cooling
Collimator: 5mm,3mm,1mm,0.5mm switch over automatically
Detector
Type: Large area SDD
Cooling way: Peltier cooling
Resolution ratio: <125eV
Sample room
Measurement situation: Atmosphere, Vacuum, Helium
Sample observation: 5 megapixel high-definition industrial camera
Computer configuration (including printer)
Host: Lenovo PC
Memory: DDR4 4G or above
Hard disk capacity: 1 TBHDD /256 GB SSD or above
OS: Windows 10
Environment setup
Temperature conditions: 10℃ - 35℃
Relative humidity: 40-70% (no dew)
Power supply: AC 220V±5V 50Hz
Rated power: 100W