Utility Requirement
Power source210-230V, 50-60Hz with electrical safety cutoff switch
Water source0.4-1 bar (5.8-14.5 psi); water supplied to fermentors should be at least 15°C below the set operating temperature
Air source0.5-2 bar, must be dry, oil-free and filtered.
SterilizationAutoclave; size of the autoclave's inner chamber must be able to accomodate vessel with condenser attached
Control unit
Control panel10.4" Color touch-screen interface
Communication portRemote software control through Ethernet, up to 16 systems per PC
Data export through USB port
Analog AUX port for system extension
Program storageUp to 59,994 process programs
Log data storageUp to 100 process monitoring data files
Cabinet materialABS front panel and painted iron housing
DimensionFootprint: W x L = 15.75" x 21.61" (400 mm x 549 mm); Height: 29.14" (740 mm)
Rated voltage220V~; 50/60 Hz, 10A, 2000W
WeightApprox. 114.6 lb (52 kg)
Aeration
Inlet gas flow-meter0,0.1-1 LPM (500 ml); 0, 0.2-2.5 LPM (1 L); 0, 1-10 LPM (3, 5 L); 0, 2-25 LPM (10 L); 0, 6-50 LPM (15, 20 L)
SpargerL-shape (500ml, 1L); Ring sparger (3L and above); Micro-sparger (C type vessels);
Center-located sparger (solid state)
Baffle316L stainless steel baffles
.5-3L vessel: fixed, unmovable
5L and above vessel: removable
Temperature
Heating1. Thermostat system: built-in heat exchanger (550W heater, water circulation pump)
2. Dry heating system : external devices (heating blanket or heating base unit)
CoolingBuilt-in water module and external water circulator (optional)
Range
- FS-V-A/ B / SA05P series: 5℃ (41℉) above coolant up to 60℃ (140℉)
- FS-V-C series (Double Jacketed): 5℃ (41℉) above coolant up to 60℃ (140℉)
- FS-V-C series (Single Wall): without temp control
- FS-V-D series : 5℃ (41℉) above coolant up to 90℃ (194℉)
ProbePlatinum RTD probe (PT-100), non autoclavable
Control modeManual or programmable 15-step PID control
Agitation
DriveRemovable top brushless motor
Speed range
a. For Pitched blade impeller: 30-300 rpm
b. For Rushton impeller: 30-1800 rpm(0.5, 1L); 30-1200 rpm(3, 5L); 30-1000 rpm(10L); 30-700 rpm(15, 20L)
c. For Broken type/Spiral type/Anchor type impellers (only for FS-V-SA05 vessel): 1 – 60 rpm
Resolution1 rpm increment
Impeller
2 impellers for 0.5-1 L vessel and 0.5-5 L Double Jacketed Vessel
3 impellers for 3 L vessel and above; for 10 L Double Jacketed Vessel
Note: customized impellers are available upon ordering
Control modeManual or programmable 15-step PID control with adjustable deadband
pH
Range0 -14 (2-12 for maximum precision)
Resolution0.01 pH
ProbeGel-filled electrode, autoclavable
Control modeManual/acid start/progrmmable 15-step PID control
DO
Range0-200%, Control range: 0-100%, adjustable
Resolution0.1%
ProbePolarographic DO sensor; autoclavable
Control mode
• 2-stage DO cascade response
a. Increase or decrease agitation speed
b. Supply external oxygen source (oxygen enrichment module required, optional device)
c. Adjust DO level using gas mixing control (gas mixing station module required, optional device)
• Substrate feeding strategy
• DO Stat with smart feeding technology
ORP (optional)
Measurement range± 2000 mV
Resolution1 mV
ProbeGel-filled electrode: autoclavable
Foam / levelProbe316L stainless steel protector with insulated PTFE tube; autoclavable, adjustable sensitivity control
Control modeFoam: on/off switch
Level: on/off switch control with wet/dry probe set up
Peristaltic pumpPump number4 built-in Watson Marlow pumps per vessel (total 8);
Total 4 external pumps expandable:
• 2 external pumps: MU-D series required (optional)
• 2 external pumps: 4-20mA or DC 0-10V analog input
** The minimum speed varies from 1-5 rpm depending on the medium viscosity. **
Motor typePrecise stepping motor; minimum speed is 1 rpm
Speed range0 - 65 rpm
Resolution1 rpm
Control modeManual or programmable 15-step feeding control; pump can be assigned for acid, base, antifoam and substrate
Exhaust
Device type:316L stainless steel condenser