Volume: 107 liters
Interior dimension: width 560 x height 480 x depth 400 mm
Exterior dimension: width 719 x height 846 x depth 591 mm
Stainless steel shelves (standard): 2
Max. number of perforated shelves: 6
Working-temperature range: +5°C above ambient temperature up to 50°C
Standard sterilisation programme: 60 minutes at 180 °C (without removing the sensors)
Setting accuracy: 0.1°C
2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
Temperature fluctuations with time at 37°C (to DIN 12880:2007-05): ± 0.1°C
Temperature variation in chamber at 37°C (to DIN 12880:2007-05): ± 0.3°C
Humidity limitation thanks to a Peltier element; when water dish is full and inserted, the Peltier element limits the value of relative humidity in the interior to 93% rh ± 2.5 %
Setting range active humidity control: 40 to 97%rh and rh-off (with option K7)
Setting accuracy humidity: 0.5% rh
Digital electronic CO2 control with dual beam NDIR system, with auto-diagnostic system and acoustic fault indication, barometric pressure compensation
Adjustment range CO2: 0 - 20%
Variation in time CO2: ± 0.2%
Setting accuracy CO2: 0.1%
ControlCOCKPIT: TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour display
Cài đặt ngôn ngữ trên bảng điều khiển ControlCOCKPIT: Đức, Anh, Pháp, Tây Ban Nha, Ba Lan, Czech, Hungary
Function SetpointWAIT: the process time does not start until the set temperature is reached
Adjustable parameters: temperature (Celsius or Fahrenheit), CO2, programme time, time zones
Interface: Ethernet LAN, USB
Programme stored in case of power failure
AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port
Additionally integrated over and undertemperature protection "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over or undertemperature, heating function is switched off in case of overtemperature, cooling function in case of undertemperature
Autodiagnostic system: integral fault diagnostics for temperature, CO2 and humidity limit control
Alarm: visual and acoustic
Large-area multi-function heating system on four sides with additional door and back heating to avoid condensation
Structured stainless steel housing, scratch-resistant, robust and durable; rear of zinc-plated steel.
Fully insulated stainless steel door with 2-point locking
Inner glass door with opening (8 mm Ø) to take gas sample
Heated inner glass door
Interior: material 1.4301 (ASTM 304), corrosion resistant
Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function HeatBALANCE: adapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to +50 %
ControlCOCKPIT with USB port for uploading programmes, reading out protocol logs, activating the User-ID function
Displaying of already logged protocol data on the ControlCOCKPIT (max 10,000 values correspond to approx. 1 week)
Ethernet interface on the rear of the appliance for reading out the protocol log and for uploading programmes and for online logging
Multiple overtemperature protection: Electronic temperature monitoring TWW/TWB (protection class 3.1) and mechanical temperature limiter TB (protection class 1) acc. to DIN 12880, AutoSAFETY automatically adjusts to the set value within a freely adjustable tolerance range. Setting individual MIN / MAX values for over/undertemperature alarm and also for all other parameters such as relative humidity, CO2
Internal data logger with a storage capacity of at least 10 years
100% AtmoSAFE
Voltage: 230V, 50/60Hz
Electrical load: aprrox. 1500 W
Ambient temperature: 10 – 35°C
Humidity: max. 70%, non-condensing
Accessories:
+ Perforated stainless steel shelf: 2 pcs
+ Stainless steel water dish: 1 pc
+ Calibration certificate at +37°C, 5 % CO2
+ Manual