Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

- Volume: 749 liters
- Interrior dimensions: 1040W x 1200H x 600D mm
- Exterior dimensions: 1124W x 1726H x 784D mm
- Stainless steel grids (standard): 2
- Max. number of grids: 14
- Max. loading per grid: 30 kg
- Max. loading of chamber: 300 kg
- Working temperature range: min. 10 above ambient temperature to 800C
- Setting accuracy: 0.10C
- 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
- Language setting: German, English, Spanish, French, Polish, Czech, Hungarian
- ControlCOCKPIT: SingleDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with high-definition TFT-colour display
- Timer: digital backwards counter with target time setting, adjustable from 1 minute to 99 days
- Function setpointWAIT: the process time dose not start until the set temperature is reached
- Calibration: three freely selectable temperature values
- Adjustable parameters: temperature (Celsius or Fahrenheit), programme time, time zones
- Forced air circulation in working chamber, fixed value 100%
- Programme stored in case of power failure
- Temperature control: adjustable electronic overtemperature monitor and mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature
- Autodiagnostic system: for fault analysis
- Door: fully insulated stainless steel door with 2-point locking, inner glass door, two leaves
- Structured stainless steel housing, scratch-resistant, robust and durable; rear of zinc-plated steel
- Internal data logger with a storage capacity of at least 10 years
- Voltage: 230 V, 50/60 Hz
- Electrical load: approx. 2000 W
- Ambient temperature: 5 – 400C
- Humidity: max. 80% rh, non-condensing
- Installation: mounted on clockable castors
Accessories:
+ Works calibration certificate at 370C (Memmert supply)
+ Stainless steel grid, 2 pcs
+ Manual