Metrel MI 3152H Standard set Multifunctional Electrical Installation Tester (50 V ~ 2500 V)
Continuity (Measuring range/ Resolution/ Accuracy)
Test Current 7 mA 2-wire:
0.00 Ω ... 19.99 Ω / 0.1 Ω / ±(5 % of r. + 3 digits)
20.0 Ω ... 1999 Ω / 1 Ω / ±(5 % of r. + 3 digits)
Test Current 200 mA 2-wire:
0.00 Ω ... 19.99 Ω / 0.01 Ω / ±(3 % of r. + 3 digits)
20.0 Ω ... 199.9 Ω / 0.1 Ω / ±(5 % of r.)
200.0 Ω ... 1999 Ω / 1 Ω / ±(5 % of r.)
Insulation Resistance (Measuring range/ Resolution/ Accuracy)
Test Voltage 50/100/250 V:
0.00 MΩ ... 19.99 MΩ / 0.01 MΩ / ±(5 % of r. + 3 digits)
20.0 MΩ ... 99.9 MΩ / 0.1 MΩ / ±(10 % of r.)
100.0 MΩ ... 199.9 MΩ / 0.1 MΩ / ±(20 % of r.)
Test Voltage 500/1000 V:
0.00 MΩ ... 19.99 MΩ / 0.01 MΩ / ±(5 % of r. + 3 digits)
20.0 MΩ ... 199.9 MΩ / 0.1 MΩ / ±(5 % of r.)
200 MΩ ... 999 MΩ / 1 MΩ / ±(10 % of r.)
Test Voltage 2500 V:
0.00 MΩ ... 19.99 MΩ / 0.01 MΩ / ±(5 % of r. + 3 digits)
20.0 MΩ ... 199.9 MΩ / 0.1 MΩ / ±(5 % of r.)
200 MΩ ... 999 MΩ / 1 MΩ / ±(10 % of r.)
1.00 GΩ ... 19.99 GΩ / 0.01 GΩ / ±(10 % of r.)
Insulation Analysis (Measuring range/ Resolution/ Accuracy)
Calculation of PI, DAR, DD:
0.01 MΩ ... 9.99 MΩ / 0.01 MΩ / ±(5 % of r. + 2 digits)
10.0 MΩ ... 100 MΩ / 0.1 MΩ / ±(5 % of r.)
RCD Test (Measuring range/ Resolution/ Accuracy)
RCD Uc (Contact Voltage):
0.00 V ... 19.99 V / 0.1 V / (-0 %/+15 %) of r. ± 10 digits
20.0 V ... 99.9 V / 0.1 V / (-0 %/+15 %) of r.
RCD (t) (Trip Time) (Measuring range/ Resolution/ Accuracy)
0.00 ms ... 40.0 ms / 0.1 ms / ±1 ms
0.0 ms ... max. time / 0.1 ms / ±3 ms
RCD I Ramp (Trip Current):
0.2×IΔN ... 2.2×IΔN / 0.05×IΔN / ±0.1×IΔN
Impedance (Measuring range/ Resolution/ Accuracy)
Zline (L-L, L-N) & Zloop (L-PE):
0.00 Ω ... 9.99 Ω / 0.01 Ω / ±(5 % of r. + 5 digits)
10.0 Ω ... 99.9 Ω / 0.1 Ω / ±(10 % of r.)
100 Ω ... 999 Ω / 1 Ω / ±(10 % of r.)
1.00 kΩ ... 9.99 kΩ / 10 Ω / ±(10 % of r.)
Voltage & Frequency (Measuring range/ Resolution/ Accuracy)
Voltage TRMS:
0 V … 550 V / 1 V / ±(2 % of r. + 2 digits)
Frequency:
0.00 Hz ... 9.99 Hz / 0.01 Hz / ±(0.2 % of r. + 1 digit)
10.0 Hz ... 499.9 Hz / 0.1 Hz / ±(0.2 % of r. + 1 digit)
Earth Resistance (Measuring range/ Resolution/ Accuracy)
3-wire Method:
0.00 Ω ... 19.99 Ω / 0.01 Ω / ±(5 % of r. + 5 digits)
20.0 Ω ... 199.9 Ω / 0.1 Ω / ±(5 % of r. + 5 digits)
200.0 Ω ... 9999 Ω / 1 Ω / ±(5 % of r. + 5 digits)
2-clamp Method:
0.00 Ω ... 19.99 Ω / 0.01 Ω / ±(10 % of r. + 10 digits)
20.0 Ω ... 30.0 Ω / 0.1 Ω / ±(20 % of r.)
30.1 Ω ... 39.9 Ω / 0.1 Ω / ±(30 % of r.)
General Data
Power Supply: 9 VDC (6 x 1.5 V battery or accu, size AA)
Overvoltage Category: 1000 V DC CAT II; 600 V CAT III; 300 V CAT IV
Protection Class: Double insulation
COM Port: BT, USB, RS232
Display: 4.3 inch Colour TFT touch screen
Dimensions (L x H x W): 230 x 103 x 115 mm
Weight: 1.3 kg
Standard set
• Instrument MI 3152H EurotestXC
• Plug commander, 1.5 m
• 2.5 kV test lead, 2 x 1.5 m
• Test lead, 3 x 1.5 m
• Power supply adapter + 6 NiMH rechargeable batteries, type AA
• Test probe, 4 pcs (blue, black, green, red)
• Crocodile clip, 4 pcs (blue, black, green, red)
• Earth set 20 m
• RS232 - PS/2 cable
• USB cable
• Soft carrying bag
• Soft carrying neck belt
• Short instruction manual