Michell Instruments Optidew 401 Chilled Mirror Hygrometer
Performance
Dew Point Measurement Accuracy: ±0.15 °C (±0.27 °F)
Repeatability: ±0.05 °C (±0.09 °F)
Sensitivity: ±0.01 °C (±0.018 °F)
Response: Stable measurement at +10 °C (+50 °F) dew point within 1 minute.
Dew-Point Sensor
1. Single Stage
Dew Point Range: –25 to +90 °C (–13 to +194 °F)
Temperature Range: –40 to +90 °C (–40 to +194 °F)
Relative Humidity Range at 23 °C: 2.25–100% RH
Material: POM sensor head, aluminum body.
2. Dual Stage
Dew Point Range: –40 to +90 °C (–40 to +194 °F)
Temperature Range: –40 to +90 °C (–40 to +194 °F)
Relative Humidity Range at 23 °C: 0.45–100% RH
Material: POM sensor head, aluminum body.
3. Harsh Environment
Dew Point Range: –40 to +120 °C (–40 to +248 °F)
Temperature Range: –40 to +120 °C (–40 to +248 °F)
Relative Humidity Range at 23 °C: 0.45–100% RH
Material: PEEK sensor head, aluminum body.
Corrosion and Saturation Protection: Active component isolation system.
Mirror Temperature Measurement: Pt1000, Class A.
Recommended Sample Flow: Ambient conditions for environmental measurements to 2 Nl/min for flowing samples.
Maximum Pressure: 2,500 kPa.
Process Connection: M36 × 1.5.
Remote PRT
Temperature Measurement Accuracy: ±0.1 °C (±0.18 °F).
Temperature Measurement: Pt100, Class A.
Cables
Cable Lengths: 0.3 m, 3 m, 5 m, 10 m, and 20 m (11.81 inches, 9.84 ft, 16.4 ft, 32.81 ft, and 65.62 ft) available. Cables can be combined.
Minimum Operating Temperature:
Standard: –25 °C (–13 °F).
High-temperature version: –40 °C (–40 °F).
Maximum Operating Temperature:
Standard: +90 °C (+194 °F).
High-temperature version: +120 °C (+248 °F).
Remote Pressure Sensor (Optional)
Pressure Measurement Accuracy: ±0.25% FS.
Pressure Measurement Range: 0–160 kPa or 0–2,500 kPa.
Process Connection: 1/8-inch NPT-M.
Control Unit
Resolution: Selectable display resolution of 1 or 2 decimal places.
Measurement Units:
Dew point: °Cdp or °Fdp.
Relative humidity: %.
Absolute humidity: g/m³.
Volume concentration: ppmV.
Mixing ratio: g/kg.
Wet-bulb temperature (Twb): °C, °F.
Water vapor pressure (wvp): Pa.
Ambient temperature: °C, °F.
Pressure-converted dew point: °C, °F.
Pressure: kPa, bara, barg, psia, psig.
Enclosure: Bench-top.
Enclosure Material: ABS.
Analog Outputs: Two 0/4–20 mA outputs (maximum load: 500 Ω).
Digital Communications:
Modbus RTU over USB (standard).
Modbus RTU over RS485 (standard).
Modbus TCP over Ethernet (optional).
Alarms:
1 process relay.
1 alarm relay.
Both are Form C, rated at 1 A, 30 VDC.
Inputs: 4–20 mA input for the pressure sensor.
Data Logging: SD card slot (standard).
Ingress Protection: IP54.
Dimensions: 220 × 175 × 118 mm (8.66 × 6.89 × 4.65 inches).
Weight:
Control unit: 1.5 kg (3.3 lb).
Sensor: 200 g (7.05 oz).
Display: 5.7-inch color touchscreen.
Environmental Conditions: –20 to +50 °C (–4 to +122 °F), up to 100% RH non-condensing; optional IP65 connector version supports 100% RH condensing conditions.
Supply Voltage: 100–240 V AC, 50–60 Hz.
Power Consumption: 30 VA maximum.