Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

Rated voltage and effective measuring range: 125V-20MΩ,250V-50MΩ,500V-100MΩ
Center scale: 0.5MΩ/1MΩ/50MΩ
Minimum measurable resistance at rated voltage: 0.125MΩ,0.25MΩ,0.5MΩ
Rated current: 1mA+20%-0%
Open circuit voltage: Rated voltage+30%-0%
Short circuit current: <2mA
Withstanding Voltage: AC 3700V, 1 minute (Between input terminal and outer case)
Storage Temperature: -10℃~60℃, 80% RH max (Without condensation)
Operating Temperature: 0℃ to 40℃, 80% RH max. (Without condensation)
Power supply: 1.5V (AM-3, R6 or "AA" size)×6
Weight: Approx. 330g (Excluding batteries)
Accuracy (DC 50V):
First effective range:0.01MΩ~5MΩ ±5%rdg、10MΩ~50MΩ ±30%rdg
Accuracy(DC 125V):
First effective range:0.02MΩ~10MΩ ±5%rdg
Second effective range:0.01MΩ~0.02MΩ 10MΩ~20MΩ ±10%rdg,10MΩ~100MΩ ±30%rdg
Accuracy (DC 250V):
First effective range:0.05MΩ~20MΩ ±5%rdg
Second effective range:0.02MΩ~0.05MΩ 20MΩ~50MΩ ±10%rdg,50MΩ~100MΩ ±30%rdg
Accuracy (DC 500V):
First effective range:0.1MΩ~50MΩ ±5%rdg
Second effective range:0.05MΩ~0.1MΩ 50MΩ~100MΩ ±10%rdg
Accuracy (DC 1000V):
First effective range:2MΩ~1000MΩ ±5%rdg
Second effective range:1MΩ~2MΩ 1000MΩ~2000MΩ ±10%rdg
AC voltage (50/60Hz):
Range/Accuracy/Input impedance/Maximum input voltage
AC 600V/±2.5% of full scale/Approx. 1.5M/AC 600V rms