Rated voltage and effective measuring range: 125V-20MΩ,250V-50MΩ,1000V-2000MΩ
Minimum measurable resistance at rated voltage: 0.125MΩ,0.25MΩ,1MΩ
Rated current: 1mA+20%-0%
Open circuit voltage: Rated voltage+30%-0%
Short circuit current: <2mA
Withstanding Voltage: AC 3700V, 1 minute (Between input terminal and outer case)
Storage Temperature: -10℃~60℃, 80% RH max (Without condensation)
Operating Temperature: 0℃ to 40℃, 80% RH max. (Without condensation)
Power supply: 1.5V (AM-3, R6 or "AA" size)×6
Weight: Approx. 330g (Excluding batteries)
Accuracy(DC 125V):
Measuring range:0~20MΩ~OL(∞)
First effective range:0.02MΩ~10MΩ<±5%rdg
Second effective range:0.01MΩ~0.02MΩ 10MΩ~20MΩ <±10%rdg
Other range:100MΩ~OL(∞) :Accuracy is not Specified
Accuracy(DC 250V):
Measuring range:0~50MΩ~OL(∞)
First effective range:0.05MΩ~20MΩ<±5%rdg
Second effective range:0.02MΩ~0.05MΩ 20MΩ~100MΩ <±10%rdg
Other range:100MΩ~OL(∞) :Accuracy is not Specified
Accuracy(DC 500V):
Measuring range:0~100MΩ~OL(∞)
First effective range:0.1MΩ~50MΩ <±5%rdg
Second effective range:0.05MΩ~0.1MΩ 50MΩ~100MΩ <±10%rdg
Other range:100MΩ~OL(∞) :Accuracy is not Specified
Accuracy(DC 1000V):
Measuring range:0~2000MΩ~OL(∞)
First effective range:2MΩ~1000MΩ <±5%rdg
Second effective range:1MΩ~2MΩ 1000MΩ~2000MΩ <±10%rdg
Other range:100MΩ~OL(∞) :Accuracy is not Specified
AC voltage (50/60Hz):
Range/Accuracy/Input impedance/Maximum input voltage
AC 600V/±2.5% of full scale/Approx. 20M/AC 600V rms