Pulse settings-2
Pulse output voltage: 0.50kV~4.00kV ±10% 0.01kV step
Pulse width: 50ns ±15%、 100ns~1000ns ±10% 50ns step
Output voltage: 0.5~4.00kV±10% (10V step)
Polarity: + / −
Rise time: <3ns
Output impedance: 50Ω
Terminal resistance: 50Ω
Pulse repetition modes
LINE PHASE: 50Hz/60Hz injection phase angle 0~360°±10°(1°step)
synchronized with L-N of EUT supply or external CDN
VARIABLE: 10ms~999ms ±10 %
EXT TRIG:
Period:>10ms
Input signal level:TTL/open collector negative logic
Pulse width:>1ms
When LINE PHASE mode is selected and there is an input with a period of 16 to 20 ms, it is recognized as a zero-cross sync signal for external CDN.
1 SHOT:
Single pulse generation each time the 1 SHOT button is pressed.
Synchronized (phase angle set on the PHASE control) or asynchronized pulse period.
Memory storage: 5 tests
Test time: 1s ~ 999s ±10% 1s step
Coupling switch: L(+), N(-), PE/PULSE OUT *manual switch by coaxial cable
Coupling mode: common-mode / normal-mode *manual switch by short plug
EUT power capacity: Single phase AC240V / DC125V 16A (L(+), N(-), PE)
External control: RS-232C compliant optical communication
Power supply: AC100~240V 50Hz/60Hz
Operating temperature /humidity: 15~35℃ / 25~75%
Dimensions / weight: (W) 430 x (H) 349 x (D) 540mm (protrusions excluded) / approx. 23kg
HV coaxial cable connector: NMHV Noiseken custom type