Technical Specifications
Control unit equipped with measuring instruments: Mass flow meters
Gas flow meter: Range from 0 to 1020 ml/min at 23 °C, 0.1 MPa, with a precision of ±1 %
Air flow meter: Range from 0 to 120 ml/min at 23 °C, 0.1 MPa, with a precision of ±1 %
Manometer: Pressure range from 0 to 0.1 MPa
Flow control valve: For gas and air
Chronometer: To check post-combustion time and flame trigger
Thermometer: Connected to the thermocouple for flame temperature measurement
General
Dimensions: 30 × 38 × 50 cm
Supplied Accessories
Burner: According to ASTM D 5205, 10 mm diameter
Thermal sensors:
nr.1 0.5 mm thermocouple type K equipped with copper block of 9 Ø mm / 9.99 gm for 125 mm flame
nr.1 0.5 mm thermocouple type K equipped with copper block of 5.5 Ø mm / 1.75 gm for 20 mm flame
Stand: For thermocouple
Reference Standard
ASTM: D 5207