Measurement Range: 0...25 %
Accuracy: +/- 0.02 % @ 1 %, +/- 0.01 % @ 0.1 %
Output Resolution: 0.01 %
Lower Detection Limit (LDL:) 0.02 % +/- 0.01
Response Time (T90): <20 seconds, typically 12 seconds
Operating Temperature Range: -20…+50 °C (-4…+122 °F)
Sample Flow Rate (application dependent)
Flow-through/extractive: 100...500 ml/min (250 ml/min optimal) in a vented atmosphere
Direct insertion: Up to 6 m/s
Life Expectancy (application dependent): Up to 24 months
Humidity: 0...95 %rh non-condensing
Process Connection: 2” Tri-clamp / Flow-through
Shelf Life: 12 months
Calibration Interval (application dependent): 6...12 months
Electrical
Display: LCD
Output Signal: 4...20 mA
Digital Communication: RS485 (not SIL-capable)
Relay Output Options: 2 x configurable alarm relays, 1 x fault alarm relay, 1 x transistor
Power Supply: 24 V DC
Relay Contact Outputs RL1 / RL2 / Alarm
Switching power max.: 62.5 V A resp max. 30 W
Switching voltage: Um 125 V AC/110 V DC
Min. current 10 uA DC
Min. voltage 10 mV DC
According to IEC 947-5-1 resp. EN 60947
Transistor Output (DO): Switching parameters: < 28 V @ < 50 mA
Intrinsically Safe Connection
Voltage Uo = 6 V DC
Current intensity Io = 0.2 mA
Power Po = 0.3 mW
Max. outer inductivity Lo = 1000 mH
Max. outer capacity Co = 10 μF
Maximum Power Consumption: 1.5 W
Indicators: LCD and LED alerts on analyzer
Mechanical
Ingress Protection: IP20 enclosure
Housing Material: PBT
Mounting: 35 mm DIN rail
Cable Length: I.S sensor connection 5 m/10 m/15 m