Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

Size Range: 0.3 to 25 μm
Size Channels: Factory calibrated at 0.3, 0.5, 1.0, 2.5, 5.0, 10.0 μm variable binning
Flow rates / Accuracy: 1.2 LPM
Concentration Limits: 15,000,000 particles/ft3 @ 10% coincidence (per ISO 21501-4); 30,000,000 particles/ft3 @ 10% coincidence (as tested and validated1)
Aerosol Concentration Range: 0.01 to 20,000 μg/m³
Battery Run Time: >10 hours continuous operation
Light Source: Long life laser diode
Resolution: 0.00003 μg/m3
Counting Efficiency: 50% @ 0.3 μm; 100% for particles >0.45 μm per JIS
Zero Count: <1 count / 60 minutes (<1 particles / 6 ft3). No fault count subtraction. No appreciable drift.
Count Modes: Particles/second, Real-Time Meter, cumulative/differential count/m³ and count/ft³, and mass concentration
Count Alarms: 1 to 9,999,999 counts
Calibration: NIST traceable
Display: 4.3″ (10.9 cm) WQVGA (480×272) color touch screen
Optional Printer: External thermal printer available
Vacuum Source: Internal pump with automatic flow control
Filtered Exhaust: Internal HEPA filter
Number of Channels: 6
Custom Size Channels: Calibration for custom size channels available
Audible Alarm: Adjustable integrated alarm
Communication Modes: Ethernet, USB, RS485, RS232, and (optional) 802.11b/g wireless
Environmental Sensors (AQM1): Includes NDIR-CO2 sensor (0-5000 ppm ±1%, resolution 1 ppm, response rate 20 seconds), temperature (32-122°F, ±1°F / 0-50°C, ±0.5°C) and relative humidity probe (15–90% RH, ±2%)
Environmental Sensors (AQM2): Includes all sensors in the 8300-AQM1 and PID-TVOC sensor (5 ppb - 50 ppm ±1.5%, response rate <3 seconds)
Alarm: Alarms for particle counts (any size channel), sensor failure, environmental sensors, and flow rate
Standards: ISO 21501-4 and JIS B9921
Instrument Calibration: Recommended minimum once per year. Gases, temperature and humidity sensors field calibratable.
External Surface: Lightweight high-impact injection molded plastic enclosure
Dimensions (L x W x H): 6.33 x 4.25 x 12.26 in (16.1 x 10.8 x 31.1 cm) includes handle and does not include probes
Weight: 2.5 lb (1.1 kg)
Accessories: User manual + Instrument Management Software (IMS) on USB flash drive, isokinetic probe, temperature relative humidity sensor, purge filter, battery, USB cable, and power supply
Optional Accessories: Printed manual, case, spare battery, external battery charger, external printer, barbed fitting, and tubing
Data Storage: Securely stores up to 45,000 samples with particle counts, environmental data, locations, and timestamps
Sample Time: 1 second to 99 hours
Power: 110 to 240 VAC 50/60 Hz universal in-line power supply
Operating Conditions: 41-104ºF (5-40ºC) / 20% to 95% non-condensing
Storage Conditions: 32-122ºF (0-50ºC) / Up to 98% non-condensing