Size Range: 0.3 to 25 μm
Size Channels: Factory calibrated at 0.3, 0.5, 1.0, 2.5, 5.0, 10.0 μm variable binning
Flow rates: 1.2 LPM
Concentration Limits: 27,000,000 particles/ft3 @ 10% coincidence (per ISO 21501-4), 50,000,000 particles/ft3 @ 10% coincidence (as tested and validated1)
Battery Run Time: >10 hours continuous operation
Light Source: Long life laser diode
Counting Efficiency: 50% @ 0.3 μm; 100% for particles >0.45 μm per JIS
Zero Count: <1 count / 60 minutes (<1 particles / 6 ft3). No fault count subtraction.
Count Modes: Real-Time Meter and graph, cumulative/differential count/m³and count/ft³, and mass concentration (PM)
Count Alarms: 1 to 9,999,999 counts
Calibration: NIST traceable
Display: 4.3″ (10.9 cm) WQVGA (480×272) color touch screen
Printer: (Optional) External thermal printer available
Vacuum Source: Internal pump with automatic flow control
Filtered Exhaust: Internal HEPA filter
Number of Channels: 6
Custom Size Channels: Calibration for custom size channels available
Audible Alarm: Adjustable built-in alarm
Battery: Removable Li-ion (Recharge time: 4 hours within instrument, <2 hours with external battery charger)
Reports: ISO 14644-1, EU GMP Annex 1, FS 209E
Recipes: 50 user-configurable recipes
Communication Modes: Ethernet and USB. (Optional) Wireless 802.11 b/g, RS485 or RS232
Environmental Sensor: Includes temperature 32º to 122ºF (0º to 50ºC) ±1ºF (0.5ºC) and relative humidity probe 15-90% ±2%
Alarm: Alarms on counts for all particle sizes, sensor failure, environmental sensors and flow
Standards: ISO 21501-4 and JIS B9921
Calibration: Recommended minimum once per year
External Surface: High impact injection molded plastic
Dimensions (L x W x H): 5.12” x 4.25” x 12.26” (13.0 cm x 10.8 cm x 31.1 cm) includes handle and does not include probes
Weight: 2.2 lb (1.0 kg)
Accessories: Operating manual and IMS software on USB flash drive, isokinetic probe, temperature relative humidity sensor, filter, battery, data download software, USB cable, power supply and cable
Optional Accessories: Printed manual, carrying case, spare battery, external battery charger, external printer and isokinetic probes, barbed fittings, IMS-RT monitoring system
Buffer Memory: 45,000 sample records (rotating buffer) including particle count data, environmental data, locations, annotations, and times. Scrollable on screen or printout.times. Scrollable on screen or printout
Sample Locations: Up to 1,000 locations 20 characters long
Sample Time: 1 second to 99 hours
Power: 110 to 240 VAC 50/60 Hz universal in-line power supply
Operating Conditions: 41º to 104ºF (5º to 40ºC) / 20% to 95% non-condensing
Storage Conditions: 32º to 122ºF (0º to 50ºC) / Up to 98% non-condensing