- Voltage measurement: 4,0 ... 600,0 V RMS value ± 0,5 % of the measured value ± 5 digits / 0,1 V
- Voltage peaks (> 10V): 50 Hz - 60 Hz / ± 5 % ± 50 digits
- Current measurements: 4,0 ... 1500,0 A real ± 1% V. of the measured value ± 5 digits / 0,01 A
- Current peaks (> 20A): 50 Hz - 60 Hz / ± 5 % ± 50 digits
-Active power P: 10 W ... 9999 kW in 5 measurement ranges
± 1 % of the value + 20 digits (>20 V y >20 A)
± 2 % of the value + 40 digits (>20 V y >20 A)
0,1 W ... 1 kW according to the measurement range
- Apparent power S: 10 VA ... 9999 kVA in 5 measurement ranges
± 1 % of the value + 20 digits (>20 V y >20 A)
± 2 % of the value + 40 digits (>20 V y >20 A)
0,1 VA ... 1 kVA according to the measurement range
- Reactive power Q: 10 VAr ... 9999 kVAr in 5 measurements ranges
± 1 % of the value + 20 digits (>20 V y >20 A)
± 2 % of the value + 40 digits (>20 V y >20 A)
0,1 VAr ... 1 kVAr according to the measurement range
- PF power factor: 0,000 ... 1,000
± 0,04 (>20 V >20 A) / 0,001
± 0,1 (>20 V >20 A) / 0,001
- Phase angle: -180 ... +180 / ±1 / 0,1
0 ... +360 / ±1 / 0,1
- Frequency measurement (U > 50 V): 46 ... 65 Hz ±0,3 Hz / 0,1 Hz
- Active energy: 0 ... 999.999 kWh
± 1 % of the value + 20 digits
- Apparent energy: 0 ... 999.999 kVAh ± 1 % of the value + 20 digits
- Reactive energy: 0 ... 999.999 kVarh ± 1 % of the value + 20 digits
- Harmonics in the voltage (relative): 50 - 60 Hz > 80 V AC
1 - 20 th / ± 2 % / 0,1 %
21 - 50 th / ± 4 % of the value ± 2 % / 0,1 %
- Harmonics in the voltage (absolute): 50 - 60 Hz > 80 V AC
1 - 20 th / ± 2 % / 0,1 %
21 - 50 th / ± 4 % of the value ± 0,5 % / 0,1 %
- Harmonics in the current (relative): 50 - 60 Hz > 80 V AC
1 - 20 th / ± 2 % / 0,1 %
21 - 50 th / ± 4 % of the valuew ± 2 % / 0,1 %
- Harmonics in the current (absolute): 50 - 60 Hz > 80 V AC
1 - 20 th / ± 2 % of the value ± 0,4 A / 0,1 A
21 - 50 th / ± 4 % of the value ± 0,4 A / 0,1 A
- Total harmonic distortion (THD-F): 50 - 60 Hz > 80 V y > 20 A
0,0 - 20 % / 2 % / 0,1 %
20,1 - 100 % / ± 6 % of the value ± 1 % / 0,1 %
100,1 - 999,9 % / ± 10 % of the value ± 1 % / 0,1 %
Max. clamp opening: 55 mm
Range selection: auto
Overload indication: "OL = Overload
Memory: 50.000 readings
Interface: USB
Software and data cable: included in the delivery. They can be used with Win 2000, XP, y ME
Display: graphic LCD display, 128 x 64 with backlight
Power supply: 2x 1,5 V AA batteries
Energy consumption: aprox. 10 mA
Dimensions: 271 x 112 x 46 mm
Weight: 650 g with batteries
Operating conditions: 85 % max. R.H. / -10 ... + 50 °C
Protection / Standards: CAT III 600 V / EN 61010-2-032
Delivery Scope
1 x 1-Phase-Power Meter PCE-GPA 62
2 x Isolated measuring pins with 2 m / 6.6 ft of cable each
2 x Batteries
1 x Carrying case
1 x USB cable for a PC
1 x Software (in English)
1 x User's manual