Scale Calibration Procedure

04/26/2021 13:44:25

**1\. Range of application** This technical document specifies a calibration procedure for scale (non-automatic scales class I) with a division value not gre

1. Range of application

This technical document specifies a calibration procedure for scale (non-automatic scales class I) with a division value not greater than 1 mg. Technical scale is (non-automatic scale class II) according to OIML R76.

2. Calibrating techniques

- Outside check

- Technical check

- Measurement check

3. Calibrating instruments

3.1 Standard measuring instrument (s)

- Analytical Balance:
Sets of weights E2, Class E2, Measuring range: 1 mg ÷ 500g  
Sets of weights F1, Class F1, Measuring range: 1g ÷ 500g
Kern ADB 200-4 Analytical balance (200g/0.1mg)


- Technical Scale:
Sets of weights F1, Class F1, Measuring range 1g ÷ 500g
Standard Weight F1, Class F1, Measurement weight: 1 kg, 2 kg, 5 kg, 10 kg.


- Bench Scale: Standard Weight M11, Class M1, Measuring range 0 ÷ 150kg



3.2 Auxiliary Instrument(s)
Hygro - Thermometer:

- Range of Measurement: (0 ÷ 50) °C
- Accuracy: ± 0,5 °C
- Range of Measurement: (10 ÷ 95) %RH
- Accuracy: ± 3 %RH

4. Conditions for Calibration
The calibration site must be bright, away from heat sources, wind sources, and not vibrated.
Temperature, humidity in the calibration site ensures:
Ambient temperature: (22 ± 3) ° C
Air humidity: (40 ÷ 80) %RH.
The scale should be calibrated at the site of use.

5. Prepare for Calibration
Before calibration, the following preparations must be done:
- The scale must be clean.
- Place the scale to be calibrated firmly on a surface and adjust the scale.
- Turn on the scale for at least 30 minutes or as required by the manufacturer.
- The scale to be calibrated should be thermally equilibrium with the environment in which it is calibrated. The minimum time required for this process is specified by the manufacturer.

6. Begin the calibration
6.1 Outside check
6.2 Technical Check
6.3 Measurement check
6.3.1 General requirements
6.3.2 Begin the calibration
- Test repeatability
- Check the deviation angle
- Check the correctness

7. Measurement Uncertainty Evaluation
8. General handling
- The calibrated scale is stamped, certified in paper.
- The calibration result includes standard value, indicator value, main number, together with the measurement uncertainty expanded for each weight level, coverage factor k.
- The recommended calibration cycle of scale is 12 months.

9. About EMIN's Calibration - Verification - Repair center
EMIN Vietnam’s Calibration and Verification center has been established since 2020 to provide after-sales services. The Calibration and Verification center is thoroughly equipped with technical equipment, followed by the assistance from various equipment manufacturers around the globe. At the center, calibration of specialized equipment is successfully carried out by experienced engineers in EMIN Vietnam, ensuring fast, accurate, reliable procedure to meet the most demanding customers’ requirements). Such technical capacity can only be performed by very few technology businesses in Vietnam. Meanwhile, after tireless effort, the Calibration and Verification center of EMIN Vietnam is officially certified to be accredited by the ISO 17025 standard (Jan, 2021)
If you want your instruments calibrated, please contact:
Equipment Verification - Calibration - Repair Service Center - EMIN Vietnam JSC 
Address: No 203 Nguyen Ngoc Vu Street, Trung Hoa Ward, Cau Giay District, Ha Noi City
Email: services@emin.vn - Hotline: 0968.060.220
Further information about scale calibration service can be found here.


Related News

Identifying the causes of short-term power outages using a power analyzer.
08/07/2026 10:28:40

A power quality analyzer can record the entire sequence of events—before, during, and after—a momentary power outage. Based on the captured voltage, current, and waveform data, engineers can pinpoint whether the fault originated from the utility grid or the facility's own electrical system, rather than relying on speculation based solely on external symptoms

Fire Alarm Response Procedure for Office Buildings
08/06/2026 15:51:06

When the fire alarm system is activated, the first priority is to assign a designated person to identify the source of the alarm and respond to the incident without delay. According to standard fire safety procedures, a trained emergency response team should immediately proceed to the affected area, assess the level of risk, and coordinate other personnel to follow the established fire emergency procedures.

Need a simple fix for weak or unstable voltage? Try a DHVN unit today.
08/06/2026 14:06:28

You can easily check for weak or unstable power right at home using a multimeter. After just a few minutes of measuring the voltage, you will be able to determine whether the power supply is experiencing a voltage drop or if the issue lies with the wiring, the outlet, or the appliance itself.

6 early signs of a burnt start capacitor to avoid losing millions
08/06/2026 10:34:44

"Silent capacitor burnout" refers to a phenomenon where a start capacitor gradually loses capacitance without exploding or emitting smoke, making it difficult for the user to detect the issue until the motor has sustained severe damage. The appliance continues to operate but runs sluggishly, overheats, and consumes twice the normal amount of electricity. Without periodic inspections, a component costing only a few tens of thousands of dong could force you to replace the entire motor, air conditioner compressor, or ceiling fan an expense running into the millions.

Stay Updated with Offers

Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.

By subscribing, you agree to our Terms of Service and Privacy Policy.

Quick Support

Direct access to our certified experts