Repeatability: ≤0.2%
Burette
Accuracy:
5 mL Burette: ±0.010 mL
10 mL Burette: ±0.020 mL
20 mL Burette: ±0.030 mL
Resolution: 1/480000
Potential titration module
Titration Repeatability: ≤0.2%
Range:
pH: (-5.000 ~ 20.000) pH
mV: (-2000.00 ~ 2000.00) mV
pX: (-2.000 ~ 20.000) pX
Ion: (1.000e-9 ~ 9.999e+9) μg/L, mg/L, g/L, mol/L, mmol/L, ppm, ppb
Resolution:
pH: 0.001 pH
mV: 0.01 mV
pX: 0.001 pX
Ion: Up to 4 significant digits
Accuracy:
pH: ±0.002 pH
mV: ±0.03% or ±0.2 mV
pX: ±0.002 pX (pX1), ±0.005 pX (pX2)
Ion: ±0.3%
Fluctuation: ±0.3 mV/3h
Repeatability: ≤0.1 mV
Conductometric titration module
Titration Repeatability: ≤0.2%
Range and Resolution:
0.000 μS/cm ~ 1.999 μS/cm, 0.001 μS/cm
2.00 μS/cm ~ 19.99 μS/cm, 0.01 μS/cm
20.0 μS/cm ~ 199.9 μS/cm, 0.1 μS/cm
200 μS/cm ~ 1999 μS/cm, 1 μS/cm
2.00 mS/cm ~ 19.99 mS/cm, 0.01 mS/cm
20.0 mS/cm ~ 199.9 mS/cm, 0.1 mS/cm
200 mS/cm ~ 2000 mS/cm, 1 mS/cm
Electronic Accuracy: ±0.5% FS
Accuracy: ±0.5% FS
Fluctuation: ±0.33% FS/3h
Repeatability: ≤0.3% FS
Dead-Stop titration module
Titration Repeatability: ≤0.2%
Polarization current Range: (0 ~ 0.199) μA, (0 ~ 1.999) μA, (0 ~ 19.99) μA, (0 ~ 199.9) μA
Polarization current detection Accuracy: ±2% FS
Polarization voltage Accuracy: ±2 mV
Photometric titration module
Titration Repeatability: ≤0.2%
Range and Resolution: (0.00 ~ 1500.00) mV, 0.01 mV
Wavelength: 470 nm, 520 nm, 570 nm, 620 nm, 650 nm
Wavelength Accuracy: ±10 nm
Thermometric titration module
Titration Repeatability: ≤0.7%
Range and Resolution: (0.0000 ~ 60.0000) ºC, 0.0001 ºC
Electronic Accuracy: ≤0.001 ºC
Temperature
Range: (-10.0 ~ 135.0) ºC
Resolution: 0.1 ºC
Accuracy: ±0.1 ºC
General
Power Supply: AC 80-250V, 47-63Hz
Dimensions: 240×360×400 mm
Net Weight: 4 kg