Spetification
Power Measurement: True RMS value of applied waveform
Power Rating: 0 - 500W (RMS)
Accuracy: ±(2% of reading +10mA)
Duty Cycle: 100% up to 60 seconds
Load Bank: 0 - 5115Ω
Resolution: 5Ω
Voltage (Peak): 0 - 10kV (Peak) - Closed load only
Voltage: 0 - 700V (RMS)
Current: 0 - 6000mA (RMS)
Crest Factor: 1.4 - 20 (Vpeak / VRMS) – The higher of the two peak voltage measurements is used for calculation
RF Leakage (High Frequency Leakage)
Active: From active part to earth
Passive: From plate - receptacle - to earth
Variable: See power measurement
Fixed: 2 x 200Ω Load
Output Connectors
Remote Foot Switch Control (CUT): 2 x 4mm - yellow, single relay contact
Remote Foot Switch Control (COAG): 2 x 4mm - blue, single relay contact
High Frequency Leakage and Power Measurements: Through 4mm sockets
USB: PC download / Future PC software
Oscilloscope Output: 5V/A, 100mA RF current minimum input
Low Frequency Filter:
100 Hz Filter: To avoid low-frequency disturbance or interference
RMS Bandwidth
Instrumentation Only: 30 Hz to 10 MHz (-3 dB)
With Loads: 30 Hz to 2.5 MHz (-3 dB)
Variable Loads: 0 - 5115Ω, steps @ 5Ω (1023 steps)
Accuracy: 1% of R
Load Array: Ceramic resistors (Non-inductive)
Measurement Delay: Foot switch delay selectable between 200 - 5000ms (10mSec resolution)
Return Electrode Monitoring
Range: 1 - 475Ω, steps @ 1Ω steps, Motor-driven potentiometer
Alarm Register: High and low, manual confirmation
Ranging: Manual or automatic
Isolation
Isolation: 10kV Isolation between measurement device and enclosure
Storage and Recall
Memory: Approx 5,000 records
Output: CSV and SSS format
General Specifications
Dimensions: 370 x 300 x 204mm
Weight: 10 kg
Operating Temperature: 15 °C to 40 °C
Storage Temperature: 0 °C to 50 °C
Mains Power: 120/230 VAC +10%; 48 to 66 Hz, 35 VA