PROBE TYPE: Insertion probe
FOR USE WITH: HygroLab C1, HP23-AW-A
OPERATING RANGE: -40…85 °C / 0…100 %rh
RANGE OF APPLICATION: -40…85 °C (-40…185 °F) / 0…100 %RH
ACCURACY AT 23 ±5 °C: ± 0.008 aw / 0.8 %rh / ±0.1 K
ACCURACY W. ADJUSTMENT PROFILE "STANDARD": at 23 °C / 10, 35, 80 %rh ± 0.8 %rh / ± 0.1 K
POWER SUPPLY: 3.2…5 VDC, calibrated at 3.3 VDC, current: ~4.5 mA
LONG-TERM STABILITY: <1 %rh / year
HUMIDITY SENSOR: HYGROMER® WA-1 / Pt-100, 1/3 DIN Class B
TEMPERATURE DEPENDENCE: 0.001 K/K, 0.0002 aw/K
TEMPERATURE SENSOR: PT100 1/3 DIN Class B
FILTER TYPE: Sintered steel with 5 μm pore size
RESPONSE TIME T 63: <20 s, with filter
ALLOWED WIND SPEED: 40 m/s, with filter
PSYCHROMETRIC CALCULATIONS: Dew or frost point
REPRODUCIBILITY AIRCHIP: <0.02 %rh / 0.01 K
START-UP TIME: 1,5 s
MEASUREMENT INTERVAL: 1 s
EXPLOITATION SOFTWARE: ROTRONIC HW4
PROBE ADJUSTMENT BY SOFTWARE: 1 Point & multipoint %rh & °C
FIRMWARE UPDATE: With internet and HW4 software
USER INFORMATION: from Software HW4 (devicemanager)
DATA PROCESSING VIA HW4: with interface cables
DEVICE PROTECTION BY PASSWORD: Yes
TYPE OF OUTPUT SIGNALS: 2 x 0…1 VDC
ANALOGUE OUTPUTS SCALEABLE BY THE USER: Yes
ANALOG OUTPUT SIGNAL (STANDARD): 0…1 V = 0…100 %rh; 0…1 V = -40…60 °C
DIGITAL OUTPUT SIGNAL (STANDARD): UART
SERVICE INTERFACE: UART
REVERSE VOLTAGE PROTECTION: Mechanically in the connector
MAX. CABLE LENGTH: Up to 5 m with passive cable. Up to 100 m with active cable.
PROBE MATERIAL: Stainless steel 1.4305 (V2A)
ENCLOSURE MATERIAL: POM
MATERIAL: Stainless steel DIN 1.4305 (probe), POM (handle)
IP PROTECTION RATING: IP50
FDA/GAMP COMPATIBILITY: FDA 21 CFR Part 11 and GAMP compatible
FDA / GAMP: FDA 21 CFR part 11 / GAMP5
CE / EMC CONFORM: 2007 / 108 / EG
DIMENSIONS: Ø10 x 280 mm