PROBE TYPE: Digital standard probe
OPERATING RANGE: -40..85 °C / 0…100 %rF
ACCURACY W. ADJUSTMENT PROFILE "STANDARD": ±0.8 %RH / ±0.1 K, at 10...30 °C; Adjustment at 23 °C and 10, 35, 80 %RH
POWER CONSUMPTION: 2.5…5.5 VDC
CURRENT CONSUMPTION: <0.5 mA @ 2.5…5.5 VDC
LONG-TERM STABILITY: <1 %rh / year
HUMIDITY SENSOR: Hygromer HT-1
TEMPERATURE SENSOR: Pt1000, Class B 1/3
PROBE PROTECTION: Polycarbonate plastic cage, black
FILTER CARTRIDGE: Polyethylene
FILTER TYPE: Polyethylene, 40 μm
RESPONSE TIME T 63: <15 sec.
ALLOWED WIND SPEED: 20 m/s, with filter
PSYCHROMETRIC CALCULATIONS: Dew or frost point
REPRODUCIBILITY AIRCHIP: <0.02 %rh / 0.01 K
EXPLOITATION SOFTWARE: RMS-Config / RMS Server Software / RMS Cloud
PROBE ADJUSTMENT BY SOFTWARE: 1 Point & multipoint %rh & °C
FIRMWARE UPDATE: Via RMS-Config / RMS Server Software / RMS Cloud
DIGITAL OUTPUT SIGNAL (STANDARD): UART
REVERSE VOLTAGE PROTECTION: Mechanically in the connector
MAX. CABLE LENGTH: Up to 5 m with passive cable. Up to 100 m with active cable.
HOUSING: Polycarbonate
FDA/GAMP COMPATIBILITY: FDA 21 CFR Part 11 and GAMP compatible
DIMENSIONS: Ø15 x 108 mm