Skylark ESD-2000U Electrostatic Discharge Generator (1 kV–20 kV ±5%)
Output voltage:
Contact discharge: 0.1 kV–1 kV ±10%; 1 kV–20 kV ±5%
Air discharge: 0.1 kV–1 kV ±10%; 1 kV–20 kV ±5%
Holding time: >5 s
Output polarity: Positive, negative, positive and negative cycle, positive and negative alternation, positive first then negative, negative first then positive
Trigger mode: SINGLE (single discharge), COUNT (discharge according to the preset number of times), 20 pps (fixed discharge rate of 20 times/s)
Test mode: General testing, gradually increasing voltage
Discharge mode: Contact discharge, air discharge
Discharge electrodes: Conical sharp electrode for contact discharge; spherical electrode for air discharge
Discharge interval: 0.05–99.99 s (0.05 s at 20 pps)
Number of discharges: 1–9999
Energy storage capacitance: 150 pF ±10% (replaceable)
Discharge resistance: 330 Ω ±10% (replaceable)
Contact discharge waveform: Complies with EN/IEC 61000-4-2 and GB/T 17626.2 standards
Rise time: 0.8 ns ±25%
First peak current:
7.5 A ±15% @ 2 kV
15 A ±15% @ 4 kV
22.5 A ±15% @ 6 kV
30 A ±15% @ 8 kV
Current at 30 ns:
4.0 A ±30% @ 2 kV
8.0 A ±30% @ 4 kV
12.0 A ±30% @ 6 kV
16.0 A ±30% @ 8 kV
Current at 60 ns:
2.0 A ±30% @ 2 kV
4.0 A ±30% @ 4 kV
6.0 A ±30% @ 6 kV
8.0 A ±30% @ 8 kV
Configured accessories: Electrodes (conical sharp ×1, spherical ×1), antistatic cloth ×1, grounding cable ×1, power cable ×1, 470 kΩ bleeder resistors ×2
Operating conditions: Temperature: 15°C–35°C; Humidity: 30%–75%
Input: AC 100–240 V, 50/60 Hz, 300 W
Dimensions and weight: D350 × W260 × H160 mm; 6 kg
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts