Display:
Type: 4 digit LED
Readout: Start, stop or ramp rate, oven temperature, and stored measurement data
Temperature:
Resolution: 0.1 °C
Accuracy:
±(0.45+0.0045T) where T is temperature in Centigrade
Typical: ±0.6 °C (<200 °C), ±1.0 °C (≥200 °C)
Reproducibility: 0.2 °C
Oven:
Range: 50 °C to 260 °C
Ramp rate: 0.5, 1.0, 2.0, 5.0, 10, 20 °C per minute
Heat-up time (typ.): 2.5 minutes to 100 °C, 6.5 minutes to 250 °C
Thermometer: Built in Pt RTD
Temperature control: Closed-loop PID
Safety shut off: Automatic after 30 minutes idle
Capillaries:
Dimensions: 1.4 mm to 1.8 mm outside dia., 100 mm length
Capacity: Up to 3 tubes
Fill height: 2 mm to 3 mm
Sample packing: Automatic (tube tapper)
General:
Power: 200 to 250 VAC, 50 to 60 Hz, 1/2 A, 75 W nominal
Environment: 0 °C to 30 °C, non-condensing
Humidity: <90%
Weight: 2 lbs.
Dimensions: 6.5"× 9.5"× 5.25" (WHD)