Measurement Ranges
Frequency (MHz): 50 Hz to 500 MHz
Power (W): Connector/Frequency Dependent:
With standard HN connector:
10 W to 10,000 W: 50 Hz to 2 MHz
10 W to 8,900 W: >2 MHz to 13.56 MHz
10 W to 6,200 W: >13.56 MHz to 27.12 MHz
10W to 5,000 W: >27.12 MHz to 40.68 MHz
10W to 4,100 W: >40.68 MHz to 60 MHZ
10 W to 2,200 W: >60 MHz to 200 MHz
10 W to 1,300 W: >200 MHz to 500 MHz
With optional 3-1/8” connector: 10 W to 10,000 W: 50 to 500 MHz
Connector Type: HN (female) standard, 3-1/8” optional
Basic Accuracy: 1.5 W + 0.3% Rdg
Power Requirements:
Calorimeter: 220 – 240 VAC, 20 A
Chiller: 220 – 240 VAC, 20 A
Chiller (coarse control): 208 VAC, 3-Phase, 16.5 A
AC Power Source: 208 VAC, 3-Phase, 42 A
AC Power Standard: 100 – 240 VAC, 10 A
Physical Dimensions:
Calorimeter Cabinet: 46 x 22.5 x 56 in (117 x 57 x 142 cm)
Chiller (coarse control): 29.25 x 28.5 x 62 in (74 x 72.4 x 157.5 cm)
AC Power Source: 32 x 25 x 56 in (81.3 x 63.5 x 142.24 cm)
Coolant Flow Rate: 4 gallons (15.2 Liters) per minute nominal
Common Specifications:
Drift: < 50% of uncertainty over 48 hours
Input Impedance: 50 Ω nominal
S11 Linear Magnitude: Typical < 0.025 at f < 100 MHz, and <0.1 at f >= 100 MHz
Communication: Ethernet
Cooling Fluid: 25% DOW-Therm SR-1, 75% deionized water
Operating Temperature: 68 to 86 °F (20 to 30 °C)
Storage Temperature: 14 to +122 °F (-10 to +50 °C)
Warranty: 1-year Parts & Workmanship for TEGAM-manufactured components