Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

Mô tả
Advanced Temperature Source for fast and precise thermal conditioning of components, parts, hybrids, modules, subassemblies, and printed circuit boards. Capable of low temperatures without the use of Liquid Nitrogen (LN2) or Liquid Carbon Dioxide (LCO2)
Performance
Temperature Range: -60 to +225°C (50/60Hz) No LN2 or LCO2 Required
Transition Rates
-40 ~ +125°C, <12 seconds
+125 ~ -40°C, <40 seconds
System Airflow Output: 5scfm (2.4 l/s) Fixed Flow Rate
Temperature control
Temperature Display & Resolution: +/- 0.1°C
Temperature Accuracy: 1.0°C (when calibrated against NIST standard)
DUT Temperature Control: Proprietary control algorithm enables DUT temperature to be directly controlled
DUT Sensor Ports: Thermocouples (type T & K)
Power:
115 ±10% VAC, 60Hz, 20amp (15amp available)
100 ±10% VAC, 50Hz, 20amp (15amp available)
200 ±10% VAC, 50Hz, 16amp
220 ±10% VAC, 60Hz, 16amp
230 ±10% VAC, 50Hz, 16amp
Compressed air
Clean, Dry Air (CDA): Filtered to 5 micron particulate contamination.
Oil Content: <0.1 ppm, by weight, filtered to 0.01 micron oil contaminant.
Dewpoint: <-70°C @ 6.2 BAR (90PSI)
Air Supply Pressure: 5.5 to 7.6 BAR (90 to 110 PSIG)
Total Air Flow Rate Required: 5.7 l/s (12 scfm), 2.8 l/s (6scfm) minimum
Air Supply Temperature: +20° to +25°C; +22°C nominal
Operating environment
Operating Temperature: +20°C ~ +28°C; +23°C nominal
Humidity: 0 ~ 60%; 45% nominal
Noise Level: <60dBA