Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

DESCRIPTION
The ECO Series is designed for any environment where audible noise and heat dissipation are concerns. It operates on an energy-efficient, 20 amp, 60Hz circuit with proprietary technology delivering quiet, smooth sounding operation. Additionally, the system does not require annual leak testing because its gas charge is below the CO2 equivalent limit for greenhouse gas emissions as specified by Regulation EU 517/2014. These fast, precise temperature forcing systems provide effective testing and conditioning of electronic components, boards and modules. Built with proven ThermoStream reliability and supported worldwide.
PERFORMANCE
Temperature Range - No LN2 or LCO2 Required: -80 to +225°C (50/60Hz)
Transition Rate:
-55 to +125°C, approx. 10 seconds
125 to -55°C, approx. 10 seconds
System Airflow Output: 4 to 18 scfm (1.9 to 8.5 l/s) Continuous
TEMPERATURE CONTROL
Temperature Display & Resolution: +/- 0.1°C
Temperature Accuracy: 1.0°C (when calibrated against NIST standard)
DUT Temperature Control: Proprietary control algorithm enables DUT temperature to be directly controlled
DUT Sensor Ports: Internal diode, thermocouples (T & K), RTD (100 Ohm platinum)
FACILITY REQUIREMENTS
Power: 200 - 250 VAC, 60Hz 20 amp, 1 phase
Compressed Air:
Filtered to 5 micron particulate contamination.
Oil Content: <0.1 ppm, by weight, filtered to 0.01 micron oil contaminant.
Dewpoint: <10°C @ 6.2 BAR (90PSI)
Air Supply Pressure: 6.2 to 7.6 BAR (90 to 110 PSIG)
Total Air Flow Rate Required: 7.1 to 14.2 l/s (15-30 scfm), 11.8 l/s (25 scfm) nominal
Air Supply Temperature: +20° to +25°C; +22°C nominal
Operating Environment: +20° to +28°C; +23°C nominal
Humidity: 0 to 60%; 45% nominal
Noise Level: <56 dBA
SERVICE & SAFETY
Refrigerants: HCFC and CFC-free, non-toxic, non-flammable
Over Temperature Protection: Settable high and low air temperature limits (factory set at +230°C)
Compliance: CE, RoHS, EU 517/2014