Parameter: LC, IR, Tr, Vt
Range: AUTO, HOLD
Trigger mode: INT/MAN/EXT/BUS
Sorting: High,Low,Pass with beeper alarm
Setting storage: 10 groups of status can be saved and loaded
Communication interface: RS232 GPIB(optional) SCPI command program supportable
Performance parameter: (condition, working temperature: 0°C─40°C, humidity : 90%RH,warmup time≥20 min)
Test voltage: 1.0V ─ 500V
Accuracy: ±(0.5% set value+0.2V)
Charge current:
- test voltage ≤100V,0.5mA─500mA;
- test voltage > 100V,0.5mA─Imax,
- Imax=50W/test voltage
- Accuracy: ±(3% set value+0.05mA)
Test range: LC: 0.001uA ─ 20.00mA/IR: 0.01kΩ ─ 99.99GΩ
Basic accuracy: LC: ±(0.3%+0.05uA)
Charge time upper limit: 0 ─ 999s manual
Test time: FAST: 40ms; MED : 60ms; SLOW: 120ms
Test condition: range is locked, trigger mode is EXT and the external trigger voltage displays the closing state
Limit setting: LC: 0 ─ 999.999mA/IR: 0 ─ 999.999GΩ
Sorting: Pass, Fail
Vf Rated involucra voltage: 1.0V ─500V
Charge current:
- 0.5mA ─ Imax
- Imax = 65W/Vf
- Accuracy: ± (3% set value+0.05mA)
Charge time upper limit: 5s ─ 600s manual