Number of Channels: 8
Withstand Test
Output Voltage:
- AC: 0.050 - 5.000kV,Step 0.001kV,Frequency 50Hz/60Hz ±0.1%,sinusoidal waveform
- DC: 0.050 - 6.000kV,Step 0.001kV
- Accuracy: ± (1% set value + 0.1% of full scale)
- Adjustment Rate: (1% output + 0.1% of full scale) rated power
Current Range:
- AC: Voltage≤4.000kV: 0.001mA - 120.0mA, Voltage>4.000kV:0.001mA – 100.0mA
- DC: 0.1uA - 20.00mA
- Accuracy: ± (1% of reading + 0.5% of full scale), AC Real: ± (1% of reading + 5% of total current reading + 5 digits)
Output Power: 500VA
ARC:
- AC: 1.0mA - 20.0mA,0.1mA Step
- DC: 1.0mA - 10.0mA,0.1mA Step
Insulation Resistance Test
Output Voltage: 0.050 - 5.000kV,Step 0.001kV ; Accuracy: ± (1% of set value + 0.1% of full scale)
Resistance Test Range: 0.100MΩ- 99.99GΩ; Resolution: 0.1MΩ
Measurement Accuracy:
≥500V:
1.000MΩ – 1.000GΩ, ± (3% of reading + 5 digits)
1.000GΩ – 10.00GΩ, ± (7% of reading + 5 digits)
10.00GΩ – 99.99GΩ, ± (10% of reading + 5 digits)
< 500V:
0.100MΩ – 1.000GΩ, ± (7% of reading + 5 digits)
1.000GΩ – 99.99GΩ,for reference only, no accuracy requirements
Time Setting
Rise Time: OFF,0.1s – 999.9s,Step 0.1s
Test Time: 0.1s - 999.9s,Step 0.1s
Fall Time: OFF,0.1s – 999.9s,Step 0.1s
Waiting Time: OFF,0.1s – 999.9s,Step 0.1s
Turn-to-turn Insulation Test
Output Pulse Voltage: 0.01kV – 6.000kV,0.01kV Step,± 5% set value ± 15V
Inductance Test Range: ≥10μH
Pulse Energy: up to 0.36 Joule
Waveform Sampling: Sampling rate: 12bit, Sampling speed: 200MHz, adjustable 8-level, Memory depth: 12k Byte, Sample average: 1 - 32
Low Resistance Test / △ and Y Type Resistance Test
Test Signal: (mΩ)1A,1Ω 0.5A,others ≤3V
Test Range: 0.01mΩ - 1.2MΩ
Resistance: Range: 0.01mΩ - 120.00mΩ; 0.1mΩ - 1200.0mΩ; 0.001Ω - 12.000Ω 0.01Ω - 120.00kΩ 0.1kΩ - 1200.0kΩ
Accuracy: ± 0.5% of reading +0.04% of full scale; ± 0.3% of reading +0.03% of full scale; ± 0.2% of reading +0.03% of full scale ± 0.1% of reading +0.03% of full scale ± 0.2% + 0.03% of full scale)
Inductance Test
Test Parameters: Ls,Lp,Rs,Rp,Q
Measurement Accuracy: 0.5%
Test Frequency: 100Hz,120Hz,1kHz, 10kHz, 100kHz
Test Signal Level: 1.0Vrms,10% accuracy