Testing Range Description
Impedance 50Ω
Test port connector N-type, female
Number of test port 2
Frequency range 100MHz~4GHz; (Low range option: 1~100MHz)
Frequency accuracy 5ppm
Frequency resolution 10Hz
Number of measurement points 2 ~ 10001
Measurement bandwidths 10Hz to 100kHz
Dynamic range( IFBW 10Hz) 120dB, typ.123 dB
Measurement parameters S11, S21, S12, S22
Testing Accuracy
Transmission measurement accuracy(magnitude/phase)
+5 dB to +10 dB 0.2dB/2º
-50 dB to +5 dB 0.1dB/1º
-70dB to -50dB 0.5 dB/3º
-90dB to -70dB 2.5 dB/11º
Reflection measurement accuracy(magnitude/phase)
-15dB to 0dB 0.4dB/3º
-25dB to -15dB 1.0dB/6º
-35dB to -25dB 3.0dB/20º
Trace stability
Trace noise magnitude( IF bandwidth 3kHz) 10 mdB rms
Temperature dependence(per one degree of temperature variation) 0.02dB
External reference input SMA female; 10 MHz; 2 dBm ± 3 dB
External reference output SMA female; 10 MHz; 3 dBm ± 2 dB
Operating temperature range +5°C ~ +40°C
Storage temperature range -45°C ~ +55°C
Operating humidity 90% (25°C)
Operating atmospheric pressure 84 to 106.7 kPa
Calibration interval 3 years
Power supply 110/220 ± 22 V (AC), 50 Hz
Power consumption 18W
Dimensions(W × H × D) mm 180 ×50×290
Weight 2.3 kg