3568 : No data output
3568-03 : BCD output, TTL level
3568-04 : BCD output, Open collector
3568-05 : RS-232C
Measurement ranges 300m-ohm、3, 30, 300, 3k and 30k-ohm、 6ranges
Accuracy Resistance measurement: ±(0.08% of rdg. + 3digit) typ.
Temperature measurement: ±(0.2% of rdg. + 0.2°)
Measurement current From DC 100mA to 10μA
Max. Applied voltage 100V, AC, DC, for all ranges (10V DC at temperature measuring range)
Sampling time 4/sec. (SLOW mode) 20/sec. (FAST mode)
Response speed 500msec. (SLOW mode) 100msec. (FAST mode)
Measurement method 4-terminal method (resistance measurement)
Temperature measuring range Pt 100 ohm : from -19.9 to 199.9 degree C
Power supply LR6 (AA) Alkaline batteries × 6, AC adaptor (100/ 200V +/- 10%)
Weight Approx. 1 kg
Dimensions 206 (W) x 65 (H) x 169 (D) mm
Accessories Kelvin clip , AC adaptor, LR6(AA)Alkaline batteries × 6