Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

- Voltage Measurement
Nominal input voltage ( kV ): 1
Maximum input voltage ( kV ): 1.1
Nominal output voltage ( V ): 10
Proportional uncertainty ( k=2, ppm ): 30
- Auxiliary Performance
Input resistance ( MΩ ): 10
Output resistance ( kΩ ): 100
Temperature coefficient ( ppm/K ): 2
Voltage influence factor ( ppb/V ): 10
- General Specifications
Power supply: ——
Warm up time: 30 minutes
Temperature performance: Operating temperature: 5°C ~ 40°C; Storage temperature: -10°C to 55°C
Note: When stored in a harsh temperature environment, the resistance value may change temporarily and be restored to the original resistance value within the following 30 days.
Humidity performance:
Working humidity :(30%-70%) R·H, no condensation
Storage humidity :(20%-80%) R·H, no condensation
Connection mode: The input connection adopts high pressure socket, and the output connection adopts a pair of ф 4 mm female socket.
Size (mm) (W * D * H) (excluding terminals): 180x180x120
Weight ( kg ): 2