Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

- Withstand voltage
Output voltage (AC ): Voltage range: 0.050kV—5.000kV
Voltage waveform: Sine wave
Distortion degree: < 3%
Operating Frequency: 50/60Hz
Frequency accuracy:±1%
Output power: 50VA(5.000kV 10mA)
Voltage regulation: ±(1.0%+50V) (Rated power)
Output voltage (DC ): Voltage range: 0.050 kV—6.000kV
Signal source: 600Hz
Output power: 30VA(6.000kV 5mA)
Voltage regulation: ±(1.0% +100V) (Rated power)
Voltage resolution: 1V
Voltage output accuracy: ±(1.0%Set value+5V) (No load)
Voltage test accuracy: ±(1.0%Reading+5V)
- ARC Detector
Measuring range: AC: Off, level 1 ~ 9
DC: Off, level 1 ~ 9
- Comparator
Discrimination mode: Window comparison mode
Current lower limit ON: pass when I lower limit < Ix < I upper limit;
When Ix ≤ I lower limit or Ix ≥ I upper limit, fail. (I lower limit < I upper limit)
Current lower limit OFF: when Ix < I upper limit, pass; When Ix ≥ I upper limit, fail.
The discrimination method of insulation resistance is the same as above
Current upper limit setting: AC: 0.001mA – 10mA
DC: 0.1uA – 5mA
Current lower limit setting (LOWER OFF): AC: 0.001mA – 10mA
DC: 0.1uA –5mA
Resistance upper limit setting: OFF, 0.1MΩ - 10GΩ
Resistance lower limit setting: 0.1MΩ– 10GΩ
Discriminant output: PASS/FAIL LCD and LED display respectively and sound alarm
Interface & protocol: USB host、 HANDLER、 RS232C, Optional RS485, SCPI protocol
Upper computer software: Yes
Power: AC 110V/220V±10% 50~60Hz
Display mode: 4.3 inch LCD screen
Product size(W×H×D): 295mm×117.6mm×465mm
Product net weight: 8.3kg