CH4:
Measurement range: 0 – 10 000 ppmv
Accuracy: ±4 ppm
±5% of reading
Repeatability: 10 ppm
5% of reading
C2H6:
Measurement range: 0 – 10 000 ppmv
Accuracy: ±10 ppm
±5 % of reading
Repeatability: 10 ppm
5% of reading
C2H4:
Measurement range: 0 – 10 000 ppmv
Accuracy: ±4 ppm
±5 % of reading
Repeatability: 10 ppm
5% of reading
C2H2:
Measurement range: 0 – 5000 ppmv
Accuracy: ±0.5 ppm
±5% of reading
Repeatability: 1 ppm
5% of reading
CO:
Measurement range: 0 – 10 000 ppmv
Accuracy: ±4 ppm
±5% of reading
Repeatability: 10 ppm
5% of reading
CO2:
Measurement range: 0 – 10 000 ppmv
Accuracy: ±4 ppm
±5% of reading
Repeatability: 10 ppm
5% of reading
H2:
Measurement range: 0 – 5000 ppmv
Accuracy: ±15 ppm
±10% of reading
Repeatability: 1 ppm
5% of reading
H2O:
Measurement range: 0 – 100 ppmw
Accuracy: ±2 ppm
±10% of reading
Repeatability: ±2 ppm
±10 % of reading
Total gas pressure:
Measurement range: 0–2000 hPa
Accuracy: ±10 hPa
±2% of reading
Repeatability: 10 hPa
5% of reading
Operating voltage:
OPTPSU1: 100–240 V AC, 50–60 Hz, ±10 %
OPTPSU2: 110–220 V DC, ±10 %
Overvoltage category: III
Maximum power consumption: 500 W
Typical power consumption at +25°C (+77°F): 100 W