Voltage kV, Current mA: 60/2-10
- Control box:
+ Rated voltage: 60kV
+ External dimension (mm): 310*250*230
+ Weight (kg): 5
- DC high voltage generator:
+ External dimension (mm): 470*260*220
+ Weight (kg): 6
Output polarity: Negative voltage polarity, zero voltage start, linear continuous adjustable
Working power supply: 50HZ, AC220V±10%
Voltage measurement accuracy: ± 0.5% ± 2 words, minimum resolution 0.1kv, measurement range (range 10% to 1.1 times of range)
Current measurement accuracy: ± 0.5% ± 2 words, minimum resolution: 0.1 μ a for anti-impact ammeter, 1 μ a for control box, measuring range (10% of range to full scale).
Ripple coefficient: ≤0.5%
Voltage stability: Random fluctuation, ≤ 0.5% when the grid voltage changes ± 10%
Operation mode: Intermittent working system, less than 30 minutes under rated load
Working conditions: Temperature: 0-40 ℃, humidity less than 90%
Storage conditions: Temperature: -10℃~40℃, humidity less than 90%
Altitude: Below 3000 meters
Note: external dimension of DC high voltage generator = diameter of equalizing cover (R) * diameter of insulating cylinder (R) * total height (H)