Temp. adjusting unit setting range: 40 to 220deg.C (inlet temperature), 0 to 60deg.C (outlet temperature)
Temperature adjusting accuracy: Inlet temperature±1deg.C
Drying air amount adjusting range: 0〜0.7m3/min
Spray air pressure adjusting range: 0〜0.3MPa
Liquid sending pump flow rate range: 0〜26 ml/min
Spray air line washing function: Spraying at the nozzle tip, manual pulse jet system
External output: Inlet temperature, outlet temperature, temperature outlet (4-20 mA)
Temperature adjusting device: PID digital temperature adjusting device
Touch panel: Blower, heater, liquid sending pump, pulse jet switch, error display
Control select switch: Inlet temperature, outlet temperature control switch (Outlet temp. control is conditional.)
Temperature sensor: K-thermocouple
Heater: 2.0 kW (at 200V) to 2.88 kW (at 240V)
Liquid sending: Fixed amount peristaltic pump
Spraying air pump: Spraying air compressor (sold separately) is used
Service outlet: For stirrer: AC100V, MAX2A
Suction blower: Bypass blower
Filter: Suction filter, exhaust filter
Recovery of solvent: -
Spray nozzle cooling mechanism: Connector: nipple x 2, O.D.:φ10.5 mm
Spray air connection diameter: Nipple diameter:φ7 mm
Spray air pressure: Bourdon tube: 0.3 MPa
Exhaust connecting diameter: φ50mm
Safety function: Inlet/outlet temperature overheat, sample feed reverse rotation mechanism, over current electric leakage breaker, nozzle connection error
External dimensions(W×D×H mm): 580×420×1125mm
Power Source: AC220V Single phase
Weight: 80kg