Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

Pressure Measurement:
+ Measurement range: Positive pressure: 0 to 200 kPa, Negative pressure: -80 to 0 kPa
+ Measurement accuracy(6 months after calibration): Positive pressure 20 to 200 kPa: ±(0.01% of reading + 0.003 kPa) 0 to 20 kPa: ±0.005 kPa; Negative pressure: ±(0.2% of reading + 0.080 kPa)
+ Measurement accuracy (1 year after calibration): Positive pressure 20 to 200 kPa: ±(0.01% of reading + 0.010 kPa)0 to 20 kPa: ±0.012 kPa; Negative pressure: ±(0.2% of reading + 0.090 kPa)
DC Current Measurement: 20 mA, 100 mA; 0.015% of reading + 3 µA at 20 mA
DC Voltage Measurement: 5 V, 50 V; 0.015% of reading + 0.5 mV at 5V
DC Current Source: 20 mA, 20 mA SIMULATE; 0.015% of setting + 3 µA
DC Voltage Source: 5V; 0.015% of setting + 0.5 mV
Display: Dot matrix LCD (320x240 dots)
Backlight: LED
Display refresh rate: Approx. 300ms (3 times/sec)
Warm-up time: Approx. 5 minutes
Power supply: Six alkaline AA batteries
Battery life: Approx. 35 hours when measuring current with the 24 V loop power supply OFF and approx 10 hours with the 24 V loop power supply ON
Auto power-off: Approx. 60 minutes (the function can be disabled)
Insulation resistance: 100 MΩ or more (500 VDC) between the input terminal and case and between the input port and case
Withstand voltage: 500 VAC for 1 minute between the input terminal and case and between the input port and case
Dimensions: Approx. 264 (W) x 188 (H) x 96 (D) mm, excluding protrusions
Weight: Approx. 2kg (including batteries)
Operating temperature / humidity ranges: -10 to 50°C and 20 to 80%RH (no condensation)
Storage temperature / humidity ranges: -20 to 60°C and 20 to 80%RH (no condensation)
Accessories: A set of 1.7m long black and red lead wires with alligator clips for generation and measurement, six alkaline AA batteries, R1 1/4” – 1/8” NPT female thread x 1, ferrite core x 2, R 1/4” – 1/4” NPT female thread x 1, accessory case, instruction manual (CD), startup guide, shoulder strap