Ainuo AN8711P(F) Digital Power analyzer
Manufacturer: Ainuo Model: AN8711P(F) - Request a Quotation
- Contact
See more: Equipment Calibration-Inspection & Repair Service
Call for the best price
Hanoi city: (024) 35.381.269
Danang city: (023) 63.747.711
Bac Ninh city: (0222) 730.39.68
Hai Phong city: (0225) 730.03.89
HCM city: (028) 38.119.636
Dong Nai: 0345.689.536
Channel: 1
Measurement parameters: U,I,P,λ,f
Voltage rated range: Crest Factor 3:60.00V / 300.0V / 600.0V
Current rated range: Crest Factor 3: 400.0mA / 4.000A / 20.00A
Voltage/current accuracy range: 40mA / 400mA / 4A : (1%~110%) × range, 40.00A : (1%~100%) × range
Power factor range: ±(0.001 ~1.000)
Voltage/current measurement accuracy: 45Hz≤f≤66Hz:±(0.1% × reading+0.1% ×range)
Active power measurement accuracy: PF=1:45Hz≤f≤66Hz:±(0.1% × reading+0.1% × range); 0<PF<1:Add tanΦ×0.2% × reading to PF=1 accuracy; PF=0:Add 0.2% × reading to PF=1 accuracy
Frequency measurement accuracy: DC,45Hz≤f≤66Hz, ±0.1% × reading
Alarming function: 10 groups, U, I, P, λ with Hi, Lo and door limit
Interface: RS232(Option RS485, GPIB)、Remote(PASS / FAIL)
Dimension: 213(W)×88(H)×386(D) mm
Weight: 3 kg
Product Detail
Introduction
AN8721P(F)/AN8711P(F) AC/DC power meter adopts ARM7 series controller as the core,using direct sampling and intellignet indentification technology,widely used for single-phase electrical equipments test,such as voltage,current,power,power factor,frequency,electric energy,time etc,Highlighted LED 3-window display ,simple and fast operation,serial communication,over limit alarm,voltage/current ratio setings and other functions,prcise and durable.
Main features
* Beautiful appearance,compact dimensions,low weight;
* Fast measurement speed,up to 200ms refresh rate;
* Voltage/current ratio can be set to expand measuring range;
* Sound & light alarm for voltage/current/power over limit,satisfy fast identification of defected products on production line;
* Minimum current measurement range is low to 0.4mA,meet the test requirement of standby power consumption;
* Wide AC/DC range,suitable for various load;
* RS-232/RS-485 port to realize automatic testing and data transmission;
* Optional I/O prot to realize remote control signal iput and alarm signal output
Functions
* Parameter sttings:voltage/current ratio,integral threshold current,electric energy accrmulating time,voltage upper/lower limti,current upper/lower limit,power upper/lower limits,alarm delay time,address,baud rate,buzzer On/Off,calculation mode and other parameters can be set,and preset parameters will be memored;
* Over limit alarm;sound & light alarm as certain parameter is out if the preset range,giving sound & light alarm signal automatically and output the digital signal(optional);
* Swithc of range:manual set to range,or power meter will automatically switch to a proper rangge under Auto range mode;
* Accumulating Time:electric energy accumulating time can be set;
* Engrgy accumulation:display the accumulated electric energy within the preset time;
* Ratio settings:voltage/current ratios can be set based on the actual measurement and the ratio of primary and secondary of external transformer,easy expansion;
* Serial communication:test results can be collected via RS-232 or RS-485(optional)on computer.
Model |
AN8721P(F) |
AN8711P(F) |
Channel |
1 |
1 |
Measurement parameters |
U,I,P,Q,S,λ,Φ,f,Wh,Time,Pavg,Pint |
U,I,P,λ,f |
Voltage rated range |
Crest Factor 3:60.00V / 300.0V / 600.0V |
Crest Factor 3:60.00V / 300.0V / 600.0V |
Current rated range |
Crest Factor 3:40.00mA / 400.0mA / 4.000A / 40.00A |
Crest Factor 3:400.0mA / 4.000A / 20.00A |
Voltage/current accuracy range |
40mA / 400mA / 4A : (1%~110%) × range, 40.00A : (1%~100%) × range |
40mA / 400mA / 4A : (1%~110%) × range, 40.00A : (1%~100%) × range |
Power factor range |
±(0.001 ~1.000) |
±(0.001 ~1.000) |
Voltage/current measurement accuracy |
DC:±(0.2% × reading+0.2% × range) 45Hz≤f≤66Hz:±(0.1% × reading+0.1% × range) |
45Hz≤f≤66Hz:±(0.1% × reading+0.1% ×range) |
Active power measurement accuracy |
DC:±(0.3% × reading+0.2% × range) PF=1: 45Hz≤f≤66Hz:±(0.1% × reading+0.1% × range) 0<PF<1:Add tanΦ×0.2% × reading to PF=1 accuracy PF=0:Add 0.2% ×reading to PF=1 accuracy |
PF=1: 45Hz≤f≤66Hz:±(0.1% × reading+0.1% × range) 0<PF<1:Add tanΦ×0.2% × reading to PF=1 accuracy PF=0:Add 0.2% × reading to PF=1 accuracy |
Power resolution |
0.01W |
0.01W |
Frequency measurement range |
DC,45Hz≤f≤66Hz |
45Hz≤f≤66Hz |
Frequency measurement accuracy |
±0.1% × reading (Voltage amplitudeshould be not less than 20% of voltage range) |
±0.1% × reading (Voltage amplitudeshould be not less than 20% of voltage range) |
Harmonic measurement |
None |
None |
Filter function |
None |
None |
Energy measurement range |
0~999999MWh |
None |
Energy measurement accuracy |
±1% × reading (up to 1.0000kWh) |
None |
Voltage/current ratio |
1~2000 |
1~2000 |
Display updating intervals |
0.2 / 0.5/ 1 s |
0.2 / 0.5/ 1 s |
Alarming function |
10 groups, U, I, P, λ with Hi, Lo and door limit |
10 groups, U, I, P, λ with Hi, Lo and door limit |
Interface |
RS232(Option RS485, GPIB)、Remote(PASS / FAIL) |
RS232(Option RS485, GPIB)、Remote(PASS / FAIL) |
Dimension |
213(W)×88(H)×386(D) mm |
213(W)×88(H)×386(D) mm |
Bracket height |
14 mm |
14 mm |
Weight |
About 3 kg |
About 3 kg |
- Quality Engagement
- Easy change and return
- Delivery Avaliable
- Favorable payment