Gastec 163 Quick-measuring Detector tubes Ethylene oxide C2H4O (0.05 ~ 3.0%)
Manufacturer: Gastec Model: 163 - Contact
See more: Equipment Calibration-Inspection & Repair Service
Call for the best price
Hanoi city: (024) 35.381.269
Danang city: (023) 63.747.711
Bac Ninh city: (0222) 730.39.68
Hai Phong city: (0225) 730.03.89
HCM city: (028) 38.119.636
Dong Nai: 0932.160.940
Indicator: 0.1 ~ 3.0%
Measuring Range: 0.05 ~ 3.0%, Number of Pump Strokes 1
Sampling Time: 45 seconds per pump stroke
Colour Change: Orange → Green
Reaction Principle: C2H4O+ Cr6+ + H2SO4 → Cr3+
Coefficient of Variation: 10% (for 0.05 to 1.0 %), 5% (for 1.0 to 3.0 %)
Nr. of tests: 10
Pump: GV-100S/110S
Measuring Range | (0.05 ) to 3.0% |
Number of Pump Strokes | 1 |
Correction Factor | 1 |
Sampling Time | 45 seconds per pump stroke |
Detecting Limit | 0.02% (n = 1) |
Colour Change | Orange → Green |
Reaction Principle | C2H4O+ Cr6+ + H2SO4 → Cr3+ |
Coefficient of Variation | 10% (for 0.05 to 1.0 %), 5% (for 1.0 to 3.0 %) |
Shelf Life | 3 Years |
Corrections for temperature & humidity | Unnecessary |
Store the tubes at cool and dark place. |
Substance | Concentration | Interference | Change colour by itself to |
Aromatic HCs | + | Green | |
Alcohols, Esters, Ethers | + | Green |
Substance | Correction Factor | No. of pump strokes | Measuring Range |
1,4-Dioxane | Factor: 2.0 | 1 | 0.1 to 6.0% |
Propylene oxide | Factor: 1.3 | 1 | 0.065 to 3.9% |
Calibration gas generation | Static gas dilution method |
TLV-TWA | TLV-STEL | Explosive range |
1 ppm(2014) | - | 3.0 to 100% |