Gastec 21 Quick-measuring Detector tubes Carbonyl sulphide COS (5~200ppm)
Manufacturer: Gastec Model: 21 - Request a Quotation
- Contact
See more: Equipment Calibration-Inspection & Repair Service
Call for the best price
Hanoi city: (024) 35.381.269
Danang city: (023) 63.747.711
Bac Ninh city: (0222) 730.39.68
Hai Phong city: (0225) 730.03.89
HCM city: (028) 38.119.636
Dong Nai: 0345.689.536
Indicator: 10 to 100 ppm
Measuring Range:
+ 5 to 10 ppm, Number of Pump Strokes 2
+ 10 to 100 ppm, Number of Pump Strokes 1
+ 100 to 200 ppm, Number of Pump Strokes 1/2
Sampling Time: 3 minute per pump stroke
Colour Change: Blue → Yellow
Reaction Principle:
- COS + I2O5 + H2SO4 → SO2 + CO2
- SO2 + BaCl2 + H2O → BaSO3+ 2HCl
- HCl + Base → Chloride
Coefficient of Variation: 10% (for 10 to 30 ppm), 5% (for 30 to 100 ppm)
Nr. of tests: 5
Pump: GV-100S/110S
Measuring Range |
|
|||
Number of Pump Strokes |
|
|||
Correction Factor |
|
|||
Sampling Time | 3 minutes per pump stroke | |||
Detecting Limit | 1 ppm (n=2) | |||
Colour Change | Blue → Yellow | |||
Reaction Principle | COS + I2O5 + H2SO4 → SO2 + CO2 SO2 + BaCl2 + H2O → BaSO3+ 2HCl HCl + Base → Chloride |
|||
Coefficient of Variation | 10% (for 10 to 30 ppm), 5% (for 30 to 100 ppm) | |||
Shelf Life | 2 Years | |||
Corrections for temperature & humidity | Unnecessary | |||
Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below. |
Substance | Concentration | Interference | Change colour by itself to |
Carbon disulphide | + | Yellow | |
sulphur dioxide | + | Yellow | |
Butane | 5000 ppm | No | |
Propane | 5000 ppm | No |
Calibration gas generation | High pressure gas cylinder method |
TLV-TWA | TLV-STEL | Explosive range |
5ppm(2014) | - | 12 to 19% |