For full functionality of this site it is necessary to enable JavaScript.

Gastec 31B Quick-measuring Detector tubes Chlorine Oxygen O2 (3~24%)

Indicator: 6 to 24%
Measuring Range: 
+ 3 to 6%, Number of Pump Strokes 1, 1 minute
+ 6 to 24%, Number of Pump Strokes 1/2, 30 seconds
Colour Change: Black → White
Reaction Principle: O2 + 4TiCl3 + 6H2O → 4TiO2 + 12HCl
Coefficient of Variation: 5% (for 6 to 24 %)
Nr. of tests: 5
Pump: GV-100S/110S

Details

Catalog

Manual

Performance
Measuring Range
3 to 6% 6 to 24%
Number of Pump Strokes
1 1/2
Correction Factor
1/2 1
Sampling Time
1 minute 30 seconds
Detecting Limit 2% (n=1)
Colour Change Black → White
Reaction Principle O2 + 4TiCl3 + 6H2O → 4TiO2 + 12HCl
Coefficient of Variation 5% (for 6 to 24 %)
Shelf Life 3 Years
Corrections for temperature & humidity Unnecessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Carbon monoxide   No No
Carbon dioxide 40% + No

Note:
If more than 40% carbon dioxide is coexisted, the tube reading will be 1% higher than the true concentration. (At 21% oxygen concentration tube 31B indicates 22%.)

 

  • Quality Engagement
  • Easy change and return
  • Delivery Avaliable
  • Favorable payment

Apply your mail to get promotion information