Gastec 91LL Quick-measuring Detector tubes Formaldehyde HCHO (0.05-1ppm)
Manufacturer: Gastec Model: 91LL Origin: Japan - Contact
- In stock: 1 Unit
See more: Equipment Calibration-Inspection & Repair Service
Call for the best price
Hanoi city: (024) 35.381.269
Danang city: (023) 63.747.711
Bac Ninh city: (0222) 730.39.68
Hai Phong city: (0225) 730.03.89
HCM city: (028) 38.119.636
Dong Nai: 0932.160.940
Indicator: 0.05-1ppm
Measuring Range: 0.05-1ppm, Number of Pump Strokes 5
Sampling Time: 1.5 minute per pump stroke
Colour Change: Yellowish Brown → Reddish Brown
Reaction Principle: 3HCHO + (NH2OH)3H3PO4 → H3PO4, H3PO4 + Base → Phosphate
Coefficient of Variation: 10% (for 0.05 to 0.3 ppm), 5% (for 0.5 to 1.0 ppm)
Nr. of tests: 10
Pump: GV-100S/110S
Measuring Range | 0.05 to 1.0 ppm |
Number of Pump Strokes | 5 |
Correction Factor | 1 |
Sampling Time | 1.5 minutes per pump stroke |
Detecting Limit | 0.03 ppm (n=5) |
Colour Change | Yellowish Brown → Reddish Brown |
Reaction Principle | 3HCHO + (NH2OH)3H3PO4 → H3PO4 H3PO4 + Base → Phosphate |
Coefficient of Variation | 10% (for 0.05 to 0.3 ppm), 5% (for 0.5 to 1.0 ppm) |
Shelf Life | 1 Year |
Corrections for temperature & humidity | Temperature correction is necessary |
Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below. |
Substance | Concentration | Interference | Change colour by itself to |
Ammonia | ![]() |
Yellow (Unclear demarcation) |
|
Nitrogen dioxide | No | Pale pink | |
Aldehydes | 1/3 time | + (Unclear demarcation) |
Reddish brown |
Ketones | 1:1 | + (Unclear demarcation) |
Reddish brown |
Toluene | 10 ppm | No | No |
Calibration gas generation | Diffusion tube method |
TLV-TWA | TLV-STEL | Explosive range |
- | C 0.3 ppm | 7.0 to 73% |