Gastec CH3CHO 92L Quick-measuring Detector tubes Acetaldehyde CH3CHO (1~20ppm)
Manufacturer: Gastec Model: CH3CHO 92L Origin: Japan - Contact
See more: Equipment Calibration-Inspection & Repair Service
Call for the best price
Hanoi city: (024) 35.381.269
Danang city: (023) 63.747.711
Bac Ninh city: (0222) 730.39.68
Hai Phong city: (0225) 730.03.89
HCM city: (028) 38.119.636
Dong Nai: 0345.689.536
Indicator: 1~20ppm
Measuring Range: 1~20ppm
Sampling Time: 2 minutes per pump stroke
Colour Change: Yellow → Brown
Reaction Principle: 3CH3CHO + (NH2OH)3H3PO4 → H3PO4, H3PO4 + Base → Phosphate
Coefficient of Variation: 10% (for 1 to 6 ppm), 5% (for 6 to 20 ppm)
Nr. of tests: 10
Pump: GV-100S/110S
Measuring Range |
|
|
Number of Pump Strokes |
|
|
Correction Factor |
|
|
Sampling Time | 2 minutes per pump stroke | |
Detecting Limit | 0.2 ppm (n=1) | |
Colour Change | Yellow → Brown | |
Reaction Principle | 3CH3CHO + (NH2OH)3H3PO4 → H3PO4 H3PO4 + Base → Phosphate |
|
Coefficient of Variation | 10% (for 1 to 6 ppm), 5% (for 6 to 20 ppm) | |
Shelf Life | 2 Years | |
Corrections for temperature & humidity | Temperature correction is necessary | |
Substance | Concentration | Interference | Change colour by itself to |
Ammonia | ![]() |
No | No |
Acetic acid | ![]() |
+ | Brown |
Acids | + | Brown | |
Aldehydes | + | Brown | |
Ketones | + | Brown |
Calibration gas generation | Diffusion tube method |
TLV-TWA | TLV-STEL | Explosive range |
- | C 25 ppm | 4 to 60% |