Gastec 1M Quick-measuring Detector tubes Carbon monoxide CO (0.05~4.0%)
Manufacturer: Gastec Model: 1M - Contact
See more: Equipment Calibration-Inspection & Repair Service
Call for the best price
Hanoi city: (024) 35.381.269
Danang city: (023) 63.747.711
Bac Ninh city: (0222) 730.39.68
Hai Phong city: (0225) 730.03.89
HCM city: (028) 38.119.636
Dong Nai: 0932.160.940
Indicator: 0.2 ~5.0%
Measuring Range:
+ 0.05 ~ 0.1%, Number of Pump Strokes 2
+ 0.1 ~ 2.0%, Number of Pump Strokes 1
+ 2.0 ~4.0%, Number of Pump Strokes 1/2
Sampling Time: 1 minute per pump stroke
Colour Change: White → Pale brown
Reaction Principle: 5CO + I2O5 + H2S2O7 → I2
Coefficient of Variation: 10% ( 0.1 ~ 0.5%), 5% (0.5 ~ 2 %)
Nr. of tests: 10
Pump: GV-100S/110S
| Measuring Range |
|
|||
| Number of Pump Strokes |
|
|||
| Correction Factor |
|
|||
| Sampling Time | 1 minute per pump stroke | |||
| Detecting Limit | 0.01% (n=2) | |||
| Colour Change | White → Pale brown | |||
| Reaction Principle | 5CO + I2O5 + H2S2O7 → I2 | |||
| Coefficient of Variation | 10% (for 0.1 to 0.5 %), 5% (for 0.5 to 2 %) | |||
| Shelf Life | 3 Years | |||
| Corrections for temperature & humidity | Unnecessary | |||
| Store the tubes at cool and dark place. | ||||
| Substance | Concentration | Interference | Change colour by itself to |
| Acetylene | + | Blackish brown | |
| Olefins | + | Pale brown | |
| Butane | + | Pale brown (whole layer) | |
| Propane | + | Pale brown (whole layer) | |
| Hexane | + | Pale brown |
| Substance | Correction Factor | No. of pump strokes | Measuring Range |
| Gasoline | 1 | 1 | 0.1 to 2% |
| Calibration gas generation | High pressure gas cylinder method |
| TLV-TWA | TLV-STEL | Explosive range |
| 25 ppm(2014) | - | 12.5 to 74% |
- Quality Engagement
- Easy change and return
- Delivery Avaliable