For full functionality of this site it is necessary to enable JavaScript.

KONICA MINOLTA CM-17d Spectrophotometer

Illumination/viewing system: di:8°,de:8° (diffuse illumination: 8° viewing), SCI (specular component included) / SCE (specular component excluded) switchable

Integrating sphere: Ø40 mm

Detector: Dual 32-element silicon photodiode arrays

Spectral separation device: Planar diffraction grating

Wavelength range: 400 nm to 700 nm

Measurement wavelength pitch: 10 nm

Half bandwidth: Approx. 10 nm

Reflectance range: 0 to 175%; Resolution: 0.01%

Light source: Pulsed xenon lamp (with UV cut filter)

Measurement time: Approx. 0.7 s (Measurement mode: SCI or SCE, from pressing trigger button to measurement completion)

Minimum measurement interval: Approx. 1.5 s (Measurement mode: SCI or SCE)

Battery performance: Approx. 2,000 measurements (approx. 1,000 measurements when using Optional WLAN/Bluetooth module) when measurements are taken at 10-second intervals at 23°C with the dedicated lithium battery, without using camera viewfinder

Measurement area/ Illumination area: MAV:Ø8 mm/Ø11 mm SAV:Ø3 mm/Ø6 mm

*Can be changed by replacing the target mask and switching the lens position

Repeatability: Standard deviation within ΔE*ab 0.02

(When a white calibration plate is measured 30 times at 5-second intervals after white calibration under Konica Minolta standard conditions)

Inter-instrument agreement: Within ΔE*ab 0.12

(Based on average for 12 BCRA Series II color tiles; MAV SCI; compared to values measured with a master body under Konica Minolta standard conditions)

Display: 2.7-inch TFT color LCD with reversible portrait viewing mode

Internal performance check*1: WAA (Wavelength Analysis & Adjustment) Technology

Interface: USB 2.0; WLAN (IEEE 802.11 b/g/n)/Bluetooth(Ver.4.1, SPP-compatible.) Optional WLAN/Bluetooth module required*2 *3

Camera viewfinder function: Using internal camera; Images can be shown on display

Observer: 2° Standard Observer, 10° Standard Observer

Illuminant:  A,C,D50,D65,F2,F6,F7,F8,F10,F11,F12,ID50,ID65,LED-B1,LED-B2,LED-B3,LED-B4,LED-B5,LED-BH1,LED-RGB1,LED-V1,LED-V2,User-defined illuminant*4 (Max. 3 types) (Simultaneous evaluation with two light sources possible)

Display items: Colorimetric values/graph, color difference values/graph, spectral graph, pass/fail judgment, pseudocolor

Color spaces: L*a*b*, L*C*h, Hunter Lab, Yxy, XYZ, and color difference in these spaces; Munsell (C)

Indices: MI, WI (ASTM E313-73/ASTM E313-98);YI (ASTM E313-73, ASTM D1925);ISO brightness (ISO2470);WI/Tint (CIE);Tristimulus Strength;Opacity; Grey scale (ISO 105-A05);gloss value; User index*5, Blackness (My) (ISO18314-3/DIN55979);Jetness (Mc) (ISO18314-3);Undertone (dM) (ISO18314-3)

Color difference equations: ΔE*ab (CIE1976) ; ΔE*94 (CIE1994); ΔE00 (CIEDE2000); CMC (l:c); Hunter ΔE; DIN99o; FMC-2;ΔE*94 (Special)*6

Data memory: 1,000 target data + 5,000 sample data

Power:

AC power supply: USB Type-C AC adapter (Power Delivery compatible, 15 W or more)

Battery: Lithium-ion battery (removable)

USB charging: USB bus power (with lithium-ion battery installed)

Charging time: Approx. 3.5 h (rapid charge) / Approx. 6 h(standard)

Size: Approx. 79(W)×230(H)×128(D) mm

Weight: Approx.700 g (Lithium-ion battery included)

Operating temperature/ humidity range: Temperature: 5 to 40°C; Relative humidity: 80% or less (at 35°C) with no condensation

Storage temperature/ humidity range: Temperature: 0 to 45°C; Relative humidity: 80% or less (at 35°C) with no condensation

Details