Voltage
Range: 100 mV; 1 V; 10 V; 100 V; 1000 V
Source Range: ±122.400 mV; ±1.22400 V; ±12.2400 V; ±122.00 V; ±1224.00 V
Accuracy (1year): 60+ 4 μV; 55+ 15 μV; 55+ 150 μV; 55+ 1.5 mV; 55+ 15 mV
Max. Output: 12 mA or more; 120 mA; 30 mA; 10 mA
Current
Range: 100 µA; 1 mA; 10 mA; 100 mA; 1 A; 10A; 30 A
Source Range: ±122.400 μA; ±1.22400 mA; ±12.2400 mA; ±122.400 mA; ±1.22400 A; ±12.2400 A; 0 to +36.720 A
Accuracy (1year): 150 + 20 nA; 70 + 30 nA; 70 + 300 nA; 90 + 3 μA; 350 + 70 μA; 380 + 1.2 mA; 540 + 1.8 mA
Max. Output: 30 V; 10 V; 2 V; 1.5 V
Temperature
Range: Pt100
Source Range: –200.0 to 850.0˚C
Accuracy (1year): ±0.12˚C
Max. Output: 0.1 to 2 mA
Resistance
Range: 400 Ω
Source Range: 1.00 to 400.00 Ω
Accuracy (1year): 75 + 0.015 Ω
Max. Output: 0.1 to 2 mA
Sweep
Target: Voltage/Current/Temperature/Resistance
Speed: Approx. 8/16/32/64 sec. selectable during 0 to 100%, 100 to 0% of setting
Output divider
Target: Voltage/Current/Temperature/ Resistance
Denominator: m 4 to 15
Numerator:n 0 to 15 (n ≤ m)
Deviation Target
Target: Voltage/Current/Temperature/Resistance
Variable range: ±20.00%
Operation: Two dials
Resolution of the first dial: 0.2% of (MAX − MIN)
Resolution of the second dial: 0.01% of (MAX − MIN)
Deviation: preset OFF/0/2%/5%
Transient response time
Voltage/Current generation: Approx. 500 ms (except for 1000 V range), approx. 3 s (1000 V range) (No load, Time to reach 0.02% of final value)
RTD/Resistance generation: Within 0.1 ms (Time constant at changing current)
Warm-up time: Approx. 30 minutes
Operating environment: Temperature: 5°C to 40°C; Humidity: 20% RH to 80% RH
Storage environment: Temperature: –15°C to 60°C; Humidity: 20% RH to 80% RH
Operating Height : 2000 m or less
Operating Attitude: Horizon
Rated supply voltage: 100 VAC to 120 VAC, 200 VAC to 240 VAC
Permitted supply voltage range: 90 VAC to 132 VAC, 180 VAC to 264 VAC
Rated supply frequency: 50 Hz/60 Hz
Permitted power supply frequency range: 48 Hz to 63 Hz
Maximum power consumption: 200 VA
Withstand voltage: 1500 VAC for 1 minute between the power supply and case.
Dimensions: 426 (W) × 177 (H) × 400 (D) mm
Weight: Approx. 13 kg