What is the maximum density for storing medicines in a refrigerator to ensure proper air circulation in compliance with GSP standards?

2026년 11월 18일 10시 49분 07초

Many pharmacies and clinics make use of every single centimeter of space inside their refrigerators to store stock. As a result, the cooling unit runs continuously, yet the temperature inside remains uneven with some areas exceeding 8°C, while others drop below 2°C, causing items to freeze

Pharmaceutical storage · GSP compliant

What Is the Maximum Medicine Load in a Pharmaceutical Refrigerator for Proper Cold Air Circulation Under GSP?

QUICK ANSWER

50 – 70%Share of the usable volume that should be filled, according to WHO guidance
5 – 7.5 cmMinimum distance from medicine boxes to the walls, ceiling, floor, and door, according to the US CDC
2 – 4 cmGap between boxes or trays so cold air can reach every surface

Sources: WHO and US CDC practice guidance on storing vaccines and pharmaceuticals.

Many pharmacies and clinics use every centimeter of their refrigerator to hold stock. The result is a unit that runs constantly while the temperature inside stays uneven: some spots climb above 8°C while others freeze below 2°C.

There is no single rigid percentage that applies to every cabinet model. International guidance sets limits based on the share of usable volume and the minimum clearances that let cold air circulate. WHO recommends filling no more than 50% to 70% of the usable volume, while the US CDC calls for 5 to 7.5 cm between medicine boxes and the cabinet surfaces. This article explains why, how to apply the numbers, and when your facility should upgrade its cabinet.

Low-temperature pharmaceutical storage

Why should you never fill a pharmaceutical refrigerator to 100% capacity?

Pharmaceutical refrigerators cool by convection. A fan pushes cold air from the evaporator through the cabinet and draws warm air back. When medicines are packed tightly, the airflow is blocked halfway, cold air cannot reach the far corners, and temperatures inside the cabinet drift apart.

1Localized temperature excursionsAreas next to the air outlet can fall below 0°C, freezing vaccines or insulin and destroying their biological activity, while hidden corners rise above 8°C.
2Compressor running nonstopThe sensor reads misleading values, so the compressor never cycles off. This shortens equipment life and wastes electricity.
3Difficulty passing GSP/GPP validationOverloading breaches the requirement for even air circulation under Circular 36/2018/TT-BYT when inspectors visit.

What maximum load density do international guidelines recommend?

The three figures below serve as the baseline when arranging medicines in the cabinet. The diagram illustrates a cabinet filled to about 60%, with free space around the walls and between boxes.

CABINET CROSS-SECTION (ILLUSTRATION)

Blue cells are medicine boxes. Dashed empty cells are space left for cold air to circulate.

50 – 70%Maximum fill ratio (WHO)Measured against the net (usable) volume, not the gross volume printed in the specifications. The remaining 30 – 50% is required so cold air can move around the boxes.
5 – 7.5 cmDistance from cabinet surfaces (CDC)Measured from the back wall, both sides, ceiling, floor, and door, to avoid dead air zones or localized freezing near the evaporator.
2 – 4 cmGap between medicine boxesCreates ventilation channels so forced cold air can reach every surface.
Regulations in Vietnam. Circular 36/2018/TT-BYT on GSP does not give a specific percentage for refrigerators. It requires storage equipment of suitable volume, with enough space to classify and arrange products by type and by batch, and with air circulating evenly. Applying WHO and CDC guidance is a practical way to meet this requirement.

Golden rules for arranging medicines in a medical refrigerator

1

Do not store medicines in the door or bottom binDoor temperatures fluctuate the most with every open and close.

2

Use ventilated trays and baskets made for the purposeAvoid sealed plastic boxes or thick cartons that block cold airflow.

3

Follow FEFO: first expired, first outArrange stock by expiry date to shorten the time the door stays open while searching.

4

Keep medicines away from the temperature sensorAvoid distorting the reading shown on the control display.

Optimizing storage space with a dedicated pharmaceutical storage cabinet

Household refrigerators have many dead corners and a freezer compartment, so their real usable volume is very small. When a healthcare facility needs to store a large quantity of medicines and still keep the load at 50 – 70%, it should invest in a dedicated pharmaceutical storage cabinet with a multi-directional forced convection fan system. Below are the capacity segments currently distributed by EMIN Vietnam.

75L – 150LClinics and small pharmacies

Compact for a moderate amount of medicine.

300L – 500LVaccination centers and mid-sized hospitals

Large enough to leave room for ventilation.

Over 1000LLarge pharmaceutical warehouses and provincial hospitals

For bulk storage facilities.

Not sure which capacity to choose? Tell EMIN how much medicine you need to store and how often the door is opened each day, and our technical team will calculate a capacity that keeps your load within the recommended range. See also EMIN's pharmaceutical storage cabinets.

References: World Health Organization (WHO) practice guidance on storing vaccines and pharmaceuticals; US Centers for Disease Control and Prevention (CDC) vaccine storage and handling materials; Circular 36/2018/TT-BYT on Good Storage Practice (GSP) for medicines.

Need a pharmaceutical storage cabinet of the right capacity?

Send us your storage needs, and EMIN will recommend a suitable model and a fast quotation.

Learn more at EMIN
© EMIN Vietnam

관련 뉴스

Purchase genuine Associated Research (ARI) equipment at EMIN, with a 5-year warranty
2026년 11월 18일 09시 19분 46초

Since September 2021, EMIN has served as the official distributor in Vietnam for brands under the Ikonix umbrella, including Associated Research. You now have a local point of contact for expert product selection advice, price quotations, and after-sales support eliminating the need to navigate the complexities of sourcing equipment independently. All genuine products come with a manufacturer’s warranty of up to 5 years, subject to the manufacturer’s terms and conditions

SMEMA standard for PCB conveyors
2026년 11월 16일 14시 00분 39초

SMEMA was developed to address this specific point; consequently, conveyors, loaders, and unloaders all include an SMEMA port in their technical specifications

Reduce scratches and drops of PCB boards during transfer between stages in an SMT production line
2026년 11월 16일 10시 42분 38초

In a well-functioning SMT line, when QC detects defects at the end of the process, the immediate assumption is often that the issue lies with the solder paste printer or the component placement machine. However, upon closer investigation, there are certain types of defects that cannot be attributed to any single specific stage. Read the article below to learn more!

혜택 소식 받아보기

대량 할인, 도매 가격 업데이트 및 신제품 소식을 이메일로 받아보세요.

구독하면 당사의 서비스 이용약관 및 개인정보 처리방침에 동의하는 것으로 간주됩니다.

빠른 지원

인증된 전문가에게 직접 연결