Test on 13MHz RFID Tag-RLB Test Method

2018年02月10日 10時17分05秒

Test on 13MHz RFID Tag-RLB Test Method Radio Frequency Identification (RFID), a wireless RF identification system, utilizes radio frequency (electromagnetic sensing, microware, etc.) to conduct the transmission, identification and swapping of wireless information.

Radio Frequency Identification (RFID), a wireless RF identification system, utilizes radio frequency (electromagnetic sensing,microware, etc.) to conduct the transmission, identification and swapping of wireless information. A complete RFID includes electronic tag (Tag), reading device (Reader) and main computer (Host). This application uses RLB to test the common 13MHz RFID Tag. This test method can also be applied for production line to rapidly compare the result with Golden Sample.

 

The Basic Understanding of the Test

The test method of RFID Tag utilizes RLB and RFID Tag fixture. After connecting TG, fixture and RF terminal, normalize the signal and place DUT at the center of fixture to obtain Tag's frequency range.

Illustration of Curve

This test method aims at obtaining the Return Loss data of DUT to find out its main operating frequency. The amplitude difference under TG Normalize line is the real test difference.

Required Equipment

GSP-9330 / GSP-9300B

Option 01: Tracking Generator

RLB-001: 1GHz Return Loss Bridge GKT-003: RLB Kit

GW RFID Tag Fixture

ADP-001 : N(M)-BNC(F) Adapter

Test Procedures

1.Connect TG, RF, RLB-001 and RFID Fixture

2.Activate GSP-9330 TG function

3. Execute TG Normalize

4.Connect DUT (RFID Tag)

5.Use Min.Peak Search to obtain the frequency of the lowest amplitude

Test Result

相关新闻

How do HRC, HRB, and HRA scales differ in hardness testing?
2026年06月15日 14時30分32秒

HRC, HRB, and HRA all belong to the Rockwell hardness system but are applied to different material groups. Understanding the differences between HRC, HRB, and HRA helps in selecting the correct testing method, accurately reading specifications, and avoiding unnecessary errors in material quality assessment.

What is the required TDS level for shrimp ponds?
2026年06月12日 14時26分27秒

What is the ideal TDS level for a shrimp tank? Many people spend a lot of time monitoring pH, temperature, and filtration systems but overlook TDS. Only when shrimp grow slowly, their colors fade, they have difficulty molting, or die off sporadically do they start checking their water source and discover that this indicator has long been out of the right range

Signs that the electrodes in your TDS meter are degraded and need replacing
2026年06月12日 11時43分51秒

TDS meters can operate reliably for many years, but the electrodes are the most subject to wear and tear. Each time a measurement is taken, the sensor surface comes into direct contact with water, dissolved minerals, and various other chemical components. This continuous process means the electrode's responsiveness gradually decreases over time.

获取优惠更新

获取专属批量折扣、批发价格更新和新产品通知,直接发送到您的邮箱。

订阅即表示您同意我们的服务条款隐私政策

快速支持

直接联系认证专家