Criteria for selecting hydrolysis particle counter

03/22/2025 08:36:49

How to choose the right particle counter for your clean room? Expensive equipment is not always the best, but it is important that it meets your needs. If you often have to monitor the environment in a factory that produces electronic components, chips or mobile phones, there will certainly be specific criteria to consider. Output connection ports, air flow, particle size to be measured, all of these factors will determine which device is suitable for your system. So what are the most important features to choose? The following article will help you understand better!

What is a particle counter?

Each type of particle counter has a specific volumetric flow rate, which directly affects the airflow velocity passing through the device. Typically, this flow rate is measured in cubic feet per minute (CFM) or liters per minute (LPM). During calibration, each counter is set to a specific flow rate, and the accuracy of particle size measurement depends on this parameter.

To meet cleanroom classification standards (ISO 14644), a particle counter collects an air sample based on volume to provide statistical data.

Standards for selecting clean room particle counters

If compliance with ISO 14644-1:2015 cleanroom standards is required, the flow rate of the particle counter will determine the sampling duration. For example, a device with a flow rate of 1 CFM needs 35 minutes to sample 1 cubic meter of air. However, a counter with a flow rate of 50 LPM reduces this time to just 20 minutes.

In cleanroom monitoring, each ISO classification defines particle concentration limits. For an ISO Class 5 cleanroom measuring 0.1 µm particles, the particle counter must be capable of detecting concentrations of at least 100,000 particles/m³ (equivalent to 2,841 particles/ft³). If monitoring the same class with a particle counter measuring 0.3 µm particles, the requirement increases to over 102,000 particles/m³ (290 particles/ft³).

One crucial point to note: the ISO standard does not specify a 0.1 µm particle concentration requirement for ISO Class 6 or higher. Therefore, using a particle counter for particles smaller than 0.1 µm is unnecessary in these cases.

Reference:

Particle Counter CEM DT-9851M

Particle Counter CEM DT-9881M

See more: EMIN Hydrolysis Particle Counter

Cleanroom Monitoring: Which Method Is Right for You?

Regular clean room monitoring

Periodic cleanroom monitoring is a mandatory requirement. This process occurs at predetermined intervals, not exceeding 60 minutes during system operation. One of the most cost-effective solutions is utilizing a Manifold system, especially if installed during the cleanroom construction phase.

A portable particle counter can be easily connected to the Manifold system at any time. This system consists of 16 or 32 sampling ports, collecting data from multiple locations and channeling it into a single particle counter in a sequential process. However, due to its cyclic sampling mechanism, there is a risk of missing sudden airborne fluctuations that the system may not capture in real time.

Continuous clean room monitoring

Continuous (online) monitoring ensures uninterrupted data collection, allowing for timely detection of any incidents. Sampling intervals can be adjusted flexibly, but higher frequency provides better accuracy.

Short sampling intervals generate large volumes of data, which can overload the system. Therefore, standard sampling intervals typically range from 1 to 10 minutes. The decision between continuous or periodic monitoring depends on budget constraints and factory infrastructure.

For detailed tracking of particle movement, the ideal approach is to assign a dedicated particle counter to each monitoring point, though this can be costly. Alternatively, if continuous sampling is not required but sufficient data is needed to assess air quality trends, a Manifold system offers a cost-effective solution. However, this method is not suitable for particles larger than 5 µm due to data transmission limitations. nếu là anh, thìa huownshg 

Connection interface of the particle particle counter

When choosing a particle counter, you should also consider the ability to connect to a central monitoring system. If the cleanroom has an Ethernet port (10 base-T or 100 base-T), prioritize a machine with network connectivity for easy integration. If the system uses serial communication, you should choose a machine that supports RS-232 or RS-485 communication standards to ensure stable data transmission.

Particle testing in laboratory environment

Laboratory particle count testing applications typically do not require compliance with ISO cleanroom standards. Instead, they focus on quantifying particles within a specific size range to determine pass or fail criteria based on the test requirements.

Selecting an appropriate particle counter depends on the type of material or sample being analyzed. Laboratories must define the critical particle size threshold (measured in µm) and the maximum allowable concentration.

High-flow-rate particle counters are often preferred due to their ability to reduce sampling time and enhance data accuracy. Since most laboratory tests target ultrafine contaminants below the micrometer scale, the range of suitable instruments is narrowed, simplifying the selection process.

Related News

The scent of new wood in a newly handed-over apartment: How to measure and address formaldehyde (HCHO) levels before moving in
11/14/2026 10:28:19

Many people refer to the pungent, acrid odor that emerges when opening a new cabinet as the "smell of new wood," assuming it will simply dissipate after a few days. In reality, however, what is being released could well be formaldehyde—a substance originating from the adhesives used in the pressed-wood panels—and our sense of smell is not a reliable enough gauge to determine the concentration levels present

Guide to Selecting a TCLP Toxicity Characteristic Leaching Procedure Shaker Configuration in Compliance with QCVN 07:2025/BNNMT and TCVN 9239:2012
11/13/2026 17:23:39

Regulation QCVN 07:2025/BNNMT (replacing QCVN 07:2009/BTNMT and QCVN 50:2013/BTNMT) specifies hazardous waste and sludge thresholds based on leachate concentrations. To ensure that analytical results for heavy metals and organic compounds are accepted during ISO/IEC 17025 or VIMCERTS assessments, laboratories must comply with the Toxicity Characteristic Leaching Procedure (TCLP) specified in TCVN 9239:2012, which is equivalent to US EPA Method 1311

Checking the Temperature of Paint and Coating Surfaces After Curing Using an Infrared Thermometer
11/13/2026 16:11:22

An infrared thermometer allows surface temperature measurement without direct contact, making it suitable for products coming out of curing ovens or surfaces that are difficult to access. However, to ensure reliable results for production quality control, it is important to understand the measurement principle and the factors that affect measurements on painted and coated surfaces.

Crowded Gyms and Yoga Studios: How to Manage Air Quality to Retain Members
11/13/2026 15:07:40

Around 6 or 7 in the evening, the gym becomes packed; the music drowns out the sound of heavy breathing, and the air conditioning runs at full blast behind closed doors. At first glance, everything seems fine: the room is cool, the lights are bright, and the equipment is state-of-the-art. Yet, after just a few sessions, regular members begin to feel a heavy-headedness and find themselves running out of breath more quickly eventually, they simply stop renewing their memberships. No one complains about the personal trainers or the exercise machines. The very thing that makes them feel stifled and drives them away is something entirely invisible: the air inside the room

Stay Updated with Offers

Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.

By subscribing, you agree to our Terms of Service and Privacy Policy.

Quick Support

Direct access to our certified experts