For full functionality of this site it is necessary to enable JavaScript.

ADEX AX-127B Digital ESR/Z checker (0.00mΩ~109.99mΩ)

Measuring range and Accuracy (at 23℃±5℃) [When the standard setting]

Test current: 100mΩ

Measuring range: 0.00mΩ~109.99mΩ

Resolution: 0.01mΩ

Measuring current: 100mA

ESR Accuracy: within ±0.5% of rdg±10digit ±100μΩ

Z Accuracy: within ±1% of rdg ±10digit ±100μΩ

Test current: 1Ω

Measuring range: 0.0mΩ~1099.9mΩ

Resolution: 0.1mΩ

Measuring current: 10mA

ESR Accuracy: within ±0.5% of rdg ±10digit

Z Accuracy: within ±1% of rdg ±10digit

Test current: 10Ω

Measuring range: 0.000mΩ~10.999mΩ

Resolution: 0.1mΩ

Measuring current: 10mA

ESR Accuracy: within ±0.5% of rdg ±10digit

Z Accuracy: within ±1% of rdg ±10digit

※The above accuracy is the value was made zero adjustment exactly after 30 minutes of warming-up.

Measuring method: 5-terminal measuring [available to contact check]

Measuring frequency: 100kHz±0.1%, SINE wave

Measuring time:

[Free running] FAST:10 times/sec. approx. SLOW:2.5 times/sec. [When the standard setting]

[Remote start] [ESR]: about (4.2+N) ×2ms [N=FAST:1~99] /SLOW: [4+power line cycle×1~9] ×2ms

(please plus 1ms., in case of making the Contact-ckeck)

[Remote start mode] [Z] about (3.5+N) ×3ms [N=FAST:1~99] /SLOW: [3.5+power line cycle×1~9] ×3ms

(please plus 1ms., in case of making the Contact-ckeck)

Comparator set range: 9999 both for HI and LO

Operation condition: [Temp.] 5℃~+40℃ [Humidity] less than 85%

Power supply: AC100V~240V selectable, 50/60Hz, about 30VA

Outer dimension: about 333 (W) ×99 (H) ×300 (D) mm (excluding protruding parts such as rubber legs, etc.)

Weight: about 4kg