Ainuo AN9637H(F) Multiple function Electrical Safety analyzer (5000V/40mA, Ground test 32A)
Manufacturer: Ainuo Model: AN9637H(F) - Request a Quotation
- Contact
See more: Equipment Calibration-Inspection & Repair Service
Call for the best price
Hanoi city: (024) 35.381.269
Danang city: (023) 63.747.711
Bac Ninh city: (0222) 730.39.68
Hai Phong city: (0225) 730.03.89
HCM city: (028) 38.119.636
Dong Nai: 0345.689.536
1) AC Hipot test
Rated output capacity: 5000V/40mA,short circuit current >200mA
Voltage range,Accuracy : 0.01~5.00kV,±1.5%
Frequency range,Accuracy: 50Hz or 60Hz,±0.1%
Current range,Accuracy: 0.10~100.00mA,±2%
2) DC Hipot test
Output voltage setting: 100~6000VDC,±2%
Voltage range,Accuracy: 0.01~6.00VDC,±1.5%
Frequency range,Accuracy: 0.0~10000uA,±2%
Current range,Accuracy: 0.1~999.9s,±0.2%
3) Insulation Resistance test
Output voltage setting: 100~2500Vdc,±2%
Resistance range, accuracy:
+ 1~9999MΩ; 100-499V,1~1000MΩ,±8%
+ 1000~2000MΩ,±12%
+ 500-2500V,1~199MΩ,±2%
+ 200~999MΩ, ± 5%
+ 1000~9999MΩ, ± 15%
4) Ground Bond test
Rated output: Max output current is 32A,max testing resistance 600mΩ,open-circuit voltage <12v br=""> Current range,accuracy: 2.0~32.0A(hoặc 64A),±2%
Voltage range,accuracy: 3.0~10.0V ,±2%
Resistance range, accuracy: 10.0~99.9~100~600mΩ;±2%
Dimension: 426x133x550mm
Weight: 30 kg
Product Detail
Main features
* Latest international standard:Comply with CCC,IEC,EN,VDE,BS,UL,JIS standard requirements,500VA WITHSTAND(AC 5KV/100mA),short-circuitoutput current exceeds 200mA,Ground resistance test current up to 64A,optional resistance or voltage mode.7 kinds of built-innetworking MD card for leakage contace current test,complying with multiple industrial standards.
★ Latest international standard: Comply with CCC, IEC, EN,VDE, BS, UL,JIS standard requirements. 500VA withstand (AC 5kV/100mA), short-circuit output current exceeds 200mA. Ground resistance test current up to 64A, optional resistance or voltage mode. 7 kinds of built-in networking MD card for leakage contact current test, complying with multiple industrial standards.
★ Intelligent test: 15 groups of test memories (group name editable), 8-step programming for each group. Product bar code reading and identification, automatically call the test group according to the model/specifications of the products tested. With U-disk to save test data timely.
★ Auto control: RS232/RS484/GPIB (four), bar code scanner interface, PLC interface, easily forming automation system.
★ Information management: optional electrical safety monitoring and control system software (ESRS), realizing test process and data management, high test efficiency, convenient for analysis and tracing.
★ Simplified operation: LCD display, shortcuts, Chinese/English.
Order information:
■AN9638H, 500VA(5000V/100mA) AC/DC withstand voltage, ground test (64A).
■AN9637H, 200VA(5000V/40mA) AC/DC withstand voltage, ground test (32A).
Spec
AC Hipot test |
||
Rated output capacity |
AN9638H(F) |
500VA(5000V/100mA),short circuit current >200mA |
AN9637H(F) |
200VA(5000V/40mA),short circuit current >200mA |
|
Output voltage setting |
Range,Accuracy |
100~5000V,±(2%×setting+5counts) |
Output frequency setting |
Range,Accuracy |
50Hz or 60Hz,±0.1%×setting |
Alarm current limits setting |
Upper-limit range,Accuracy |
0.10~100.00mA(or 40.00mA),±(2%×setting+5counts) |
Lower-limit range,Accuracy |
0.000~9.999mA,±(2%×setting+5counts) |
|
Time setting |
Range |
0.5~999.9s(0 is infinite) |
Ramp up/down time range |
0.1~999.9s |
|
Accuracy |
±(0.2%×setting+1count) |
|
Measurement |
Voltage range,Accuracy |
0.01~5.00kV,±(1.5%×reading+1count) |
Frequency range,Accuracy |
50Hz or 60Hz,±0.1%×reading |
|
Current range,Accuracy |
0.10~100.00mA(or 40.00mA),±(2%×reading+2counts) |
|
Time range,Accuracy |
0.1~999.9s,±(0.2%×reading+1count) |
|
Output voltage waveform, distortion, regulation |
Sine wave,<2%(pure resistance load),<2%×setting+5V)(from empty load to full load) |
|
Start voltage setting |
(0%~50%)×output voltage setting |
|
Current compensation setting |
0.000~100.00mA(or 40.00mA),auto、manual |
|
Arc detection |
1~9 level,9 level is optimally sensitive,0 is OFF |
|
DC Hipot test |
||
Output voltage setting |
Range,Accuracy |
100~6000VDC,±(2%×setting+5counts) |
Alarm current limits setting |
Upper-limit,Accuracy |
0~10000μA,±(2%×setting+5counts) |
lower-limit,Accuracy |
0.0~999.9μA,±(2%×setting+5counts) |
|
Time setting |
Range |
0.5~999.9s(0 is infinite) |
Ramp up time range |
0.1~999.9s |
|
Ramp down time range |
0.3~999.9s(0 is OFF) |
|
Accuracy |
±(0.2%×setting+1 count) |
|
Measurement |
Voltage range,Accuracy |
0.01~6.00VDC,±(1.5%×reading+1count) |
Frequency range,Accuracy |
0.0~10000uA,±(2%×reading+5count) |
|
Current range,Accuracy |
0.1~999.9s,±(<span style="font-size:9.0pt; font-family:"Times New Roman |
- Quality Engagement
- Easy change and return
- Delivery Avaliable
- Favorable payment