Output Rating: 5 kVA @ 20 mAAC
6 kVA @ 7.5 mADC
Maximum Limit:
AC
Range:0.00 – 20.00 mA
Resolution: 0.01 mA
DC
Range:0 – 7500 µA
Resolution: 1 µA
Accuracy: AC and DC ± (2% of setting
+ 2 counts)
Minimum Limit :
AC
Range:0.000 – 9.999 mA
Resolution: 0.001 mA
DC
Range:0.0 – 999.9 µA
Resolution: 0.1µA
Accuracy: AC and DC ± (2% of setting
+ 2 counts
Ground Continuity Current DC 0.1A ± 0.01 A, fixed
Ground Continuity
Maximum Limit
Minimum Limit
Range: 0.00 – 1.50 Ω
Resolution 0.01 Ω
Accuracy: ± (3% of setting + 0.02 Ω)
INSULATION RESISTANCE TEST MODE
Voltage Setting 30 – 1,000 VDC
Resolution: 1V
Accuracy:± (2% of setting + 5 V)
Memories: 50 steps
1500 test results
Interface USB standard
Dimensions (215 mm x 88.1 mm x 300 mm)